—#185060
Product
Artis zee/ zeego & Artis Q/ Q.zen Artis Pheno
- FDA product code
- OWB — Interventional Fluoroscopic X-Ray System
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K181407
- Affected lot / code info
- Model number: Artis zee floor- 10094135 Artis zee ceiling- 10094137 Artis zee multi-purpose- 10094139 Artis zee biplane- 10094141 Artis zeego- 10280959 Artis Pheno- 10849000 Artis Q floor- 10848280 Artis Q ceiling- 10848281 Artis Q biplane- 10848282 Artis Q zeego- 10848283 Artis Q.zen floor- 10848353 Artis Q.zen ceiling- 10848354 Artis Q.zen biplane- 10848355
Why it was recalled
When the user changes frame rates from lower frequency (e.g. 10 f/s) to higher frequency (e.g. 30 f/s) during continuous release of Fluoro "Fluoro Override", the measured Air Kerma Rate can exceed the regulatory limits.
Root cause (FDA determination)
Software design
Action the firm took
Siemens Medical Solutions USA, Inc. has initiated a Customer Advisory Notice to all affected customers via AX067/20/P on November 30, 2020. All modifications and repairs are provided to affected customers free of charge.
Recalling firm
- Firm
- Siemens Medical Solutions USA, Inc
- Address
- 40 Liberty Blvd, Malvern, Pennsylvania 19355
Distribution
- Distribution pattern
- Worldwide-US Nationwide
Timeline
- Recall initiated
- 2020-11-30
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #185060. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.