Recalls / —
—#185723
Product
Medtronic 6F Launcher Guide Catheter 100 cm JR 4.0 Model Number: LA6JR40
- FDA product code
- DQY — Catheter, Percutaneous
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- 510(k) numbers
- K021256
- Affected lot / code info
- Lot Code: 220754416, 220754417
Why it was recalled
Sterility may be compromised due to an unsealed pouch
Root cause (FDA determination)
Process control
Action the firm took
Medtronic issued 15-Jan-2021 Medical Device Correction letter via UPS 2-day to consignees stating reason for recall, health risk, and action to take: Medtronic records indicate that your facility has received one or more of the affected Launcher Guide Catheters. As a result, Medtronic requests that you immediately take the following actions: "Identify and quarantine all unused affected 6F Launcher Guide Catheters as listed in the table above. "Return/Exchange all unused affected product in your inventory to Medtronic. Contact Medtronic Customer Service at1-800-716-6700 to initiate a product return/exchange. Your local Medtronic Representative can assist you in thereturn of this product. "Please complete the enclosed Customer Confirmation Form and email to rs.cfqfca@medtronic.com. This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Questions regarding this communication, please contact your Medtronic Field Representative.
Recalling firm
- Firm
- Medtronic Vascular
- Address
- 35-37A Cherry Hill Drive, Danvers, Massachusetts 01923-2565
Distribution
- Distribution pattern
- Nationwide
Timeline
- Recall initiated
- 2021-01-15
- Terminated
- 2023-06-15
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #185723. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.