FDA Device Recalls

Recalls /

#186451

Product

Manifold rotator (Component): a. 250 2 GANG MANIFOLD ROTATOR, OFF HANDLE, Model Number MX1232MR b. 250 3 GANG MANIFOLD ROTATOR, OFF HANDLE, Model Number MX1233MR c. 550 2 GANG MANIFOLD ROTATOR, OPEN HANDLE, Model Number MX1322MR d. 550 3 GANG MANIFOLD ROTATOR, OFF HANDLE, Model Number MX1333MR. Stopcocks and manifolds are used to control the direction of IV fluid flow. component of extravascular blood-pressure Transducer.

FDA product code
DRSTransducer, Blood-Pressure, Extravascular
Device class
Class 2
Medical specialty
Cardiovascular
Affected lot / code info
a. Model Number MX1232MR, Lot Numbers: 4027757 b. Model Number MX1233MR, Lot Numbers: 4076456 c. Model Number MX1322MR, Lot Numbers: 3992363, 4011839 d. Model Number MX1333MR, Lot Numbers: 4016439

Why it was recalled

Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring devices and interventional imaging devices.

Root cause (FDA determination)

Process control

Action the firm took

The firm initiated the action by email on 03/31/2021. The Recall Notice instructs consignees to identify, and quarantine affected product in their possession. A list of affected product has been included on the Recall Notice for consignees to identify affected product. The consignees have been instructed to complete the response form, and return it to fieldactions@smiths-medical.com within 10 days of receipt. After the completed Recall Response Form has been submitted to fieldactions@smiths-medical.com, the consignee will be contacted to arrange return of any affected product in their possession. Distributors have been instructed that if they have distributed potentially affected product to their customers to immediately notify them of the Recall.

Recalling firm

Firm
Smiths Medical ASD Inc.
Address
6000 Nathan Ln N, Minneapolis, Minnesota 55442-1690

Distribution

Distribution pattern
Distribution US nationwide including Puerto Rico, Austria, Canada, Chile, Ecuador, Hong Kong, Mexico, Slovenia, Brazil, South Africa, Colombia, and Japan.

Timeline

Recall initiated
2021-03-31
Posted by FDA
2021-04-21
Terminated
2023-03-20
Status