FDA Device Recalls

Recalls /

#186864

Product

DIALYSIS ACCESS Medical convenience kit, Avid Medical Part Number: VMED004-03.

FDA product code
LROGeneral Surgery Tray
Device class
Class 2
Medical specialty
General, Plastic Surgery
Affected lot / code info
Part Number:VMED004-03; Package GTIN: 20809160238247; Lot Number: 1409970 (Expiration Date: 04/26/2022), 1412734 (Expiration Date: 10/01/2021), 1413146 (Expiration Date: 10/01/2021), and 1413395 (Expiration Date: 04/25/2022)

Why it was recalled

Medical convenience kits are being recalled due to a potential breach in the package integrity of a component, the BD ChloraPrep 3mL applicator. Under certain circumstances the applicator may grow Aspergillus penicillioides, which may result in a breach of the package integrity. A breach of package integrity may contaminate other components in AVID Medical kits.

Root cause (FDA determination)

Environmental control

Action the firm took

A Recall notification letter titled, "URGENT MEDICAL DEVICE CORRECTION" was sent to consignees on 04/09/2021 via a email. End-users were notified on 4/13/21. The letter instructs the consignee to immediately examine all inventory locations and discontinue use of all Halyard or AVID branded Procedure Trays subject to the field action. Discard all affected Procedure Trays found in inventory. In addition, if the consignees has further distributed this product, the notice instructs the consignee to please identify their customers and notify them at once of this product correction and requirements. The notification to the consignee's customers may be enhanced by including a copy of this URGENT MEDICAL DEVICE CORRECTION Notice. The consignee is also asked to please complete the attached Field Action Response Form (page 3) immediately, even if they no longer have any of the recalled lot numbers left in inventory. The completed form is to be emailed to: QualityAssurance@owens-minor.com. If the facility has identified affected product in the consignee's inventory, the recall coordinator will reach back to their facility and provide the consignee with paperwork for documenting destruction of the recalled product.

Recalling firm

Firm
Avid Medical, Inc.
Address
9000 Westmont Dr, Toano, Virginia 23168-9351

Distribution

Distribution pattern
Distributed US nationwide to TX, AZ, IL, MS, LA, MO, and TN

Timeline

Recall initiated
2021-04-09
Posted by FDA
2021-06-04
Terminated
2022-07-18
Status

Source: openFDA Device Recall endpoint. Recall record ID #186864. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.