Recalls / —
—#187102
Product
NM/CT 860 Nuclear Medicine / CT Scanners
- FDA product code
- KPS — System, Tomography, Computed, Emission
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K173816
- Affected lot / code info
- GTIN*00840682140751
Why it was recalled
GE Healthcare has become aware of a potential issue on 800 family NM/CT Scanners that could cause a shift in CT radiation exposure range of up to 5cm compared to the intended radiation exposure range of the planned scan under specific workflows. The issue occurs only on a hybrid whole-body continuous F3 protocol with Zoom < 1 where the scan range is set on the Smart Console. In some cases, this may also necessitate a re-scan of the patient, which would expose the patient to additional X-ray radiation.
Root cause (FDA determination)
Device Design
Action the firm took
GE Healthcare will send a customer notification letter describing the issue and providing safety instructions for immediate mitigation of the issue. Customers were instructed, they may continue to use the system and follow workaround instructions (zoom=1 for specific WB protocol) until the correction is provided. GE Healthcare corrective action is the development and deployment of a software update on all affected smart console IB systems at no cost to the customer. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Recalling firm
- Firm
- GE Healthcare, LLC
- Address
- 3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615
Distribution
- Distribution pattern
- Worldwide distribution - US Nationwide
Timeline
- Recall initiated
- 2021-02-07
- Posted by FDA
- 2021-05-11
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #187102. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.