FDA Device Recalls

Recalls /

#188195

Product

The system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions. These images may be obtained either withor without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories. This device may include data and image processing to produce images in a variety of transaxial and reformatted planes. Further the images can be post processed to produce additional imaging planes or analysis results. The system is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages. The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.

FDA product code
JAKSystem, X-Ray, Tomography, Computed
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K203169
Affected lot / code info
Revolution CT

Why it was recalled

When adding one or more intended scan groups names must contain only letters, numbers (0-9), or a space or the user interface will not update to confirm the intended group(s) has been added to the acquisition.

Root cause (FDA determination)

Process design

Recalling firm

Firm
GE Healthcare, LLC
Address
3000 N Grandview Blvd, Waukesha, Wisconsin 53188-1615

Distribution

Distribution pattern
US

Timeline

Recall initiated
2021-06-30
Status

Source: openFDA Device Recall endpoint. Recall record ID #188195. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.