FDA Device Recalls

Recalls /

#188883

Product

Bivona Aire-Cuf Endotracheal Tube

FDA product code
BTRTube, Tracheal (W/Wo Connector)
Device class
Class 2
Medical specialty
Anesthesiology
510(k) numbers
K792035
Affected lot / code info
SKU 25W080 Lots #'s: 3648874 and 3890459 SKU 25W085 Lot # 3387669 SKU 25W090 Lots #'s 322657 and 3342023 SKU 25W095 SKU 35W080 Lots #'s: 3237503, 3237504, 3263029, 3263030, 3267969, 3276517, 3276518, 3276519, 3301034, 3301035, 3301036, 3301479, 3307931, 3312816, 3312817, 3312818, 3312819, 3315710, 3315711, 3315712, 3346087, 3346088, 3349554, 3350510, 3350511, 3361918, 3361919, 3370120, 3370121, 3370122, 3379494, 3423620, 3433341, 3437242, 3437243, 3437244, 3437245, 3442385, 3450988, 3450989, 3451017, 3454875, 3454876, 3459488, 3459489, 3523718, 3528478, 3536391, 3536392, 3536393, 3536395, 3536398, 3536399, 3540955, 3541009, 3565654, 3572777, 3572790, 3577132, 3577134, 3577142, 3582755, 3582756, 3596290, 3604256, 3604272, 3608155, 3622333, 3622334, 3622336, 3630052, 3637293, 3656655, 3661490, 3661491, 3661492, 3664416, 3664417, 3677633, 3677634, 3677635, 3700907, 3700908, 3700909, 3700910, 3721738, 3721739, 3726299, 3730889, 3748028, 3890125, 3890127, 3890128, 3890129, 3890130, 3890131, 3890132, 3890133, 3890134, 3917895, 3917896, 3926899, 3926900, 3931090, 3933423, 3936455, 3952241, 3952242, 3952243, 3952244, 3952245, 3952246, 3955677, 3962474, 3985139, 3985140, 3985141, 3985142, 3985143, 3985144, 3985145, 3985146, 3991321, 3991322, 4026705, 4026706, 4039134, 4042021, 4042022, 4043952, 4043953, 4047982, 4060251, 4060252, 4063132, 3459490, 3463693, 3463694, 3463695, 3463696, 3472875, 3472876, 3484265, 3484266, 3484267, 3484273, 3484274, 3488047, 3504014, 3507849, 3512360, 3751160, 3751163, 3774410, 3778485, 3778493, 3782472, 3794166, 3800536, 3890117, 3890118, 3890119, 3890120, 3890121, 3890122, 3890123, 3890124, 4063133, 4063134, 4063135, 4063136, 4071990, 4071991, 4080805, 4080806, 4083993, 4089280, 4095958, 4097867, 4105268, 4105269, 3463693. SKU 35W085 Lots #'s: 3218553, 3230635, 3230636, 3238617, 3238618, 3258464, 3312935, 3346089, 3370123, 3437246, 3491405, 3507907, 3517266, 3540956, 3569173, 3590693, 3596299, 3608106, 3677631, 3748025, 3774374, 3870331, 3870332, 3890194, 3890195, 3890196, 3890197, 3901068, 3917865, 3929833, 3931093, 3948979, 3998080, 4015630, 4035792 and 4047987. SKU 35W090 Lots #'s: 3469110, 3904517,4002358, 3504012, 3929861, 4002359, 3552747, 3959226, 4029451, 3700911, 3991328 and 4060253. SKU 35W095 Lots #'s: 3317880, 3839565, 4042023, 3360929 and 4035809

Why it was recalled

Inadequate pouch seal leading to a compromised sterile barrier on products with a shaft length of 325mm.

Root cause (FDA determination)

Process control

Action the firm took

Instructions to Consignees 1. Locate and quarantine affected product in your possession. 2. Determine the number of affected devices in your possession and complete the Response Form 3. All affected product must be returned to Sedgwick for processing. A pre-paid return shipping label will be sent to you when the Response Form is returned indicating whether impacted devices are in your possession. Instructions to Distributors DISTRIBUTORS: if you have distributed potentially affected product to your customers, please immediately notify them of this Recall and instruct them to return the Response Forms to YOU for reconciliation. Please respond for each Distribution Center affected only

Recalling firm

Firm
Smiths Medical ASD Inc.
Address
6000 Nathan Ln N, Minneapolis, Minnesota 55442-1690

Distribution

Distribution pattern
Worldwide distribution - US Nationwide distribution in the states of AL, AZ, CA, CT, FL, GA, IL, KY, MA, MN, MO, NE, NJ, NY, OH, PA, SD, TN, TX, VA, WA and Hawaii. The countries of Bermuda, Canada, Germany, Denmark, Great Britain and Italy.

Timeline

Recall initiated
2021-08-09
Terminated
2023-10-13
Status

Source: openFDA Device Recall endpoint. Recall record ID #188883. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.