Recalls / —
—#189039
Product
MR MAGNETOM Systems with RT Image Suite with syngo MR VA20A, VA30A and VA31A with license SW-Key CT_RT_IMA_SUITE_SYNTHETIC_CT Material Numbers: 11060815
- FDA product code
- LNH — System, Nuclear Magnetic Resonance Imaging
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K203443
- Affected lot / code info
- Serial Numbers: 176302 175681
Why it was recalled
Potential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the treatment planning system (TPS).
Root cause (FDA determination)
Software design
Action the firm took
Siemens has initiated three (3) Customer Safety Advisory Notices (CSANs) to affected customers via SY052/21/S, CT042/21/S and MR019/21/S by email or hand delivery by the Siemens Customer Service Engineer on 8/16/21. Temporary workarounds are available for some of the software versions and are described in the CSANs. Additionally, a solution to eliminate the root cause of this problem is being distributed via SY051/21/S (for SY and CT) and MR021/21/S for MR to all affected customers free of charge. The service organization will contact you to arrange a date to perform this corrective action free of charge. Contact the service organization for an earlier appointment at 1-800-888-7436. ***UPDATE: 01/31/22 Additionally, a solution to eliminate the root cause of this problem is being distributed via CT049/21S, CT061/21/S, CT062/21/S, and CT088/21/S for CT. ***UPDATE: 03/12/22 Status Report Additional 13 S/N's added
Recalling firm
- Firm
- Siemens Medical Solutions USA, Inc
- Address
- 40 Liberty Blvd, Malvern, Pennsylvania 19355-1418
Distribution
- Distribution pattern
- US Nationwide distribution.
Timeline
- Recall initiated
- 2021-08-16
- Posted by FDA
- 2021-09-10
- Status
- —