Product

SuperCore Biopsy Instrument: REF/Product Description: 701114090/SUPERCORE BIOPSY INSTR 14GA X 9CM; 701114150/SUPERCORE BIOPSY INSTR 14GA X 15CM; 701116090/SUPERCORE BIOPSY INSTR 16GA X 9CM; 701118060/SUPERCORE BIOPSY INSTR 18GA X 6CM; 701118090/SUPERCORE BIOPSY INSTR 18GA X 9CM; 701118150/SUPERCORE BIOPSY INSTR 18GA X 15CM; 701118200/SUPERCORE BIOPSY INSTR 18GA X 20CM; 701120090/SUPERCORE BIOPSY INSTR 20GA X 9CM; 701120150/SUPERCORE BIOPSY INSTR 20GA X 15CM; 701120200/SUPERCORE BIOPSY INSTR 20GA X 20CM; 701214090/SUPERCORE BIOPSY INSTR W/COAXIAL - 14G x 9CM; 701214150/SUPERCORE BIOPSY INSTR W/COAXIAL - 14G x 15CM; 701216090/SUPERCORE BIOPSY INSTR W/COAXIAL - 16G x 9CM; 701216150/SUPERCORE BIOPSY INSTR W/COAXIAL - 16G x 15CM; 701218090/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 9CM; 701218150/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 15CM; 701218200/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 20CM; 701220090/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 9CM; 701220150/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 15CM; 701220200/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 20CM; 701114090M/SUPERCORE BIOPSY INSTR 14GA X 9CM-MAMMOTOME; 701114150M/SUPERCORE BIOPSY INSTR 14GA X 15CM-MAMMOTOME; 701116090M/SUPERCORE BIOPSY INSTR 16GA X 9CM-MAMMOTOME; SC 1409/PGI Supercore 14GA x 9CM; SC 1609/PGI Supercore 16GA x 9CM; SC 1806/PGI Supercore 18GA x 6CM;

FDA product code
KNWInstrument, Biopsy
Device class
Class 2
Medical specialty
Gastroenterology, Urology
510(k) numbers
K974814
Affected lot / code info
REF/Product Description: 701114090/SUPERCORE BIOPSY INSTR 14GA X 9CM; Lot Numbers: 11364350 11365643 11368388 11371872 11374998 11376772 11378864 11382837 11385962 11389848 701114150/SUPERCORE BIOPSY INSTR 14GA X 15CM; Lot Numbers: 11368389; 11376773; 701116090/SUPERCORE BIOPSY INSTR 16GA X 9CM; Lot Numbers: 11362716; 11363320; 11365935; 11377634; 11381340; 11384192; 11388461; 11388552; 11389139; 701118060/SUPERCORE BIOPSY INSTR 18GA X 6CM; Lot Numbers: 11361702; 11365788; 11371012; 11374235; 11376598; 11379722; 11382768; 11383890; 11388656; 11389138; 701118090/SUPERCORE BIOPSY INSTR 18GA X 9CM; Lot Numbers: 11362759; 11363229; 11364420; 11367439; 11370106; 11372917; 11376074; 11377430; 11378217; 11383193; 11383785; 11384243; 11387238; 11388864; 701118150/SUPERCORE BIOPSY INSTR 18GA X 15CM; Lot Numbers: 11370107; 11375663; 11377819; 11382602; 11385139; 11385965; 11389140; 11390503; 701118200/SUPERCORE BIOPSY INSTR 18GA X 20CM; Lot Numbers: 11372225; 11374410; 11377431; 11384193; 11388505; 701120090/SUPERCORE BIOPSY INSTR 20GA X 9CM; Lot Numbers: 11374411; 11378865; 11383280; 701120150/SUPERCORE BIOPSY INSTR 20GA X 15CM; Lot Numbers: 11366252 11367440; 11376374; 11377820; 11381387; 11385960; 701120200/SUPERCORE BIOPSY INSTR 20GA X 20CM; Lot Numbers: 11366253; 11380018; 11388893; 701214090/SUPERCORE BIOPSY INSTR W/COAXIAL - 14G x 9CM; Lot Numbers: 11369916; 11378293; 11383587; 11387065; 701214150/SUPERCORE BIOPSY INSTR W/COAXIAL - 14G x 15CM; Lot Numbers: 11377967; 11382843; 701216090/SUPERCORE BIOPSY INSTR W/COAXIAL - 16G x 9CM; Lot Numbers: 11364421; 11367443; 11374684; 11379724; 11385966; 11388551; 701216150/SUPERCORE BIOPSY INSTR W/COAXIAL - 16G x 15CM; Lot Numbers: 11382845; 11388550; 11390251; 701218090/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 9CM; Lot Numbers: 11361701; 11364247; 11367445; 11372366; 11374807; 11378687; 11383192; 11383784; 11388286; 701218150/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 15CM; Lot Numbers: 11371404; 11373084; 11377636; 11380993; 11383194; 11383195; 11387400; 11387833; 11390504; 701218200/SUPERCORE BIOPSY INSTR W/COAXIAL - 18G x 20CM; Lot Numbers: 11366315; 11372227; 11377432; 11381193; 11382998; 11384716; 701220090/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 9CM; Lot Numbers: 11367446; 11370574; 11375664; 11380589; 11381572; 11388507; 701220150/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 15CM; Lot Numbers: 11368750; 11371876; 11372228; 11376375; 11377429; 11380104; 11384717; 11387148; 701220200/SUPERCORE BIOPSY INSTR W/COAXIAL - 20G x 20CM; Lot Numbers: 11362655; 11363516; 11372322; 11378295; 11380103; 11382884; 11387584; 701114090M/SUPERCORE BIOPSY INSTR 14GA X 9CM-MAMMOTOME; Lot Numbers: 11371356; 11382375; 11389615; 701114150M/SUPERCORE BIOPSY INSTR 14GA X 15CM-MAMMOTOME; Lot Number: 11371357; 701116090M/SUPERCORE BIOPSY INSTR 16GA X 9CM-MAMMOTOME; Lot Number: 11371359; SC 1409/PGI Supercore 14GA x 9CM; Lot Numbers: 11370550; 11384881; SC 1609/PGI Supercore 16GA x 9CM; Lot Number: 11370552; SC 1806/PGI Supercore 18GA x 6CM; Lot Numbers: 11375600; 11364469;

Why it was recalled

Due to component (housing and plunger) detaching from the semi-automatic biopsy instruments.

Root cause (FDA determination)

Device Design

Action the firm took

On 09/30/2021, the firm sent an "Urgent-Product Recall Notice" letter to customers via US Postal Service Registered Mail informing them that the firm has received complaints regarding the semi-automatic Biopsy instrument coming apart during shipping or prior to use. According to the firm's investigation, the plastic housing and plunger can be separated more easily than normal for the lots manufactured during this time frame. Customers are instructed to return all affected product to the Recalling Firm's Athens TX facility using RGA#26683, attention Andrea Wieczor. Mailing address is: RGA# 26683 Argon Medical Devices, Inc. 1445 Flat Creek Road Athens, TX 75751 USA Att: Andrea Wieczor In addition, Customers are instructed to complete the response acknowledgement form included with the letter/email and return by fax, mail, or email as soon as possible. For questions. contact the Director, Post-Market Experience at email address Brian.Rogers@argonmedical.com, or contact Arbee Cummings at Arbee.Cummings@argonmedical.com or Andrea Wieczor at Andrea.Wieczor@argonmedical.com. (Updated information 11/2/2021), on 10/28/2021, the firm sent an updated "Urgent-Product Recall Notice for Additional Lots" letter via mail informing customers that they are expanding the recall to additional lots (list on the Inventory Sheet attached to recall notice). Again the Recalling Firm is instructing customer to return the affected product (listed on the Inventory Sheet).

Recalling firm

Firm
Argon Medical Devices, Inc
Address
1445 Flat Creek Rd, Athens, Texas 75751-5002

Distribution

Distribution pattern
Worldwide distribution - U.S. Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV. The countries of ALB, ARE, ARG, AUS, AUT, BEL, BGD, BRA, CAN, CHE, CHL, CHN, COL, CZE, DEU, ECU, ESP, EST, FRA, GBR, GRC, HKG, HUN, IND, IRN, ISR, ITA, JOR, JPN, MAR, MEX, MYS, NLD, NPL, NZL, PAK, PER, PHL, POL, PRT, SAU, SGP, SRB, THA, TUR, and TWN.

Timeline

Recall initiated
2021-10-04
Status

Source: openFDA Device Recall endpoint. Recall record ID #189610. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.