FDA Device Recalls

Recalls /

#189744

Product

Oncentra Brachy users using version 4.0 and above with Applicator Modeling license 170283-00

FDA product code
MUJSystem, Planning, Radiation Therapy Treatment
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K132816
Affected lot / code info
Oncentra Brachy users using version 4.0 and above with Applicator Modeling license 170283-00 GTIN Numbers: 08717213052758, 08717213038660, 08717213020610, 08717213052321, 08717213052314, 08717213052307, 08717213052291, 08717213052246, 08717213052239, 08717213051881, 08717213051843, 08717213051782, 08717213051775, 08717213051294, 08717213020610, 08717213053717, 08717213053700

Why it was recalled

Under certain circumstances, there can be a difference between the planned and delivered dose of the two ovoid channels.

Root cause (FDA determination)

Software design

Action the firm took

An Urgent Important Field Safety Notification letter (FAC-NU-0007) was issued to consignees to notify of the issue, clinical impact and recommend the following: 1) Recommended User Action in the unique scenario, which is only applicable for the Venezia" applicator: Avoid use of the combination of the two Venezia" applicator Anchor Points O1-Center and O2-Center by disabling one of them in the Applicator Library Manager. 2)Recommended User Action in the exceptional situation that may also occur for other Applicator Models: Avoid incorrect placement of the Anchor Points, especially exchanging left / right anchor point positions. 3) Recommended User Actions for both scenarios: Until an updated version of Oncentra Brachy is available, it is strongly advised: " To use additional Anchor Points during applicator modelling. This will further reduce the chance of mirroring the Applicator Model. " To cross-check the source dwell position information from the Case Explorer to be in accordance with the source dwell positions in the image views. " Use the Triangles option for the applicator skin visualization in the 3D view, and review whether the Applicator Model has a correct shading display. " Perform proper Quality Assurance for all treatment plans before treatment delivery. Consignees are required to acknowledge receipt of the notification to Elekta within 30 days.

Recalling firm

Firm
Elekta, Inc.
Address
400 Perimeter Center Ter NE, Ste 50, Atlanta, Georgia 30346-1227

Distribution

Distribution pattern
Distribution throughout USA OUS distribution including:Algeria, Argentina, Australia, Austria Bahamas, Bangladesh, Belarus, Belgium, Bolivia, Botswana, Brazil Brunei, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran (Islamic Republic of), Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kosovo, Libyan Arab Jamahiriya, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Peru, Philippines, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Saudi Arabia, Senegal, Singapore, Slovakia, South Africa, Spain, Sudan, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam

Timeline

Recall initiated
2021-10-13
Status

Source: openFDA Device Recall endpoint. Recall record ID #189744. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.