Product
Artis Q ceiling with software VD12-angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model Number: 10848281
- FDA product code
- OWB — Interventional Fluoroscopic X-Ray System
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K181407
- Affected lot / code info
- Serial Numbers: 110088 110059 110103 110035 110046 110005 110058 110056 110062 109523 109645 110067 110095 110051 110001 110020 110007 110093 110069 110075 109627 110042 110043 110048 110061 110063 110064 110032 110047 110102 110039 110040 110023 110065 110052 110018 110072 110081 110036 110054 110053 110044 110031 110016 110037 110077 109497 110079 110099 110027 110049 110066 110024 110017 110025 110100 110083 110091 110006 110055 110026 110022 110000 133501 110057 110104 110041 Expanded Recall 11/15/21: 110096 110014
Why it was recalled
If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure
Root cause (FDA determination)
Software design
Action the firm took
Siemens Medical Solutions USA, Inc. initiated a Customer Safety Advisory Notice dated 10/202/1 to customers via AX070/21/S, Siemens will update the system software via Update Instruction AX069/21/S. Letter states reason for recall, health risk and action to take: To avoid the problem, it is recommended to position the SID first and then release x-ray. Siemens will correct the software error (software VD12) via Update Instruction AX069/21/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. We do not consider it necessary to re-examine any patients in relation to the issues described above. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein
Recalling firm
- Firm
- Siemens Medical Solutions USA, Inc
- Address
- 40 Liberty Blvd, Malvern, Pennsylvania 19355-1418
Distribution
- Distribution pattern
- US Nationwide Distribution
Timeline
- Recall initiated
- 2021-10-20
- Terminated
- 2024-06-18
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #190134. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.