Recalls / —
—#194414
Product
NATURALYTE Liquid Acid 2.0K 2.5Ca- indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for acute and chronic renal failure Part Number: 08-2251-0
- FDA product code
- KPO — Dialysate Concentrate For Hemodialysis (Liquid Or Powder)
- Device class
- Class 2
- Medical specialty
- Gastroenterology, Urology
- 510(k) numbers
- K192017
- Affected lot / code info
- UDI-DI: (01) 000840861101696 (17) 240331 (10) 22CTAC106 Lot Number: 22CTAC106 Exp Date: 2024-03-31
Why it was recalled
Alternate Bottle Caps labeled with "Sealed for your Protection" on the foil not qualified for use
Root cause (FDA determination)
Under Investigation by firm
Action the firm took
Fresenius Medical Care issued Urgent Medical Device Recall letter (FA-2022-11-T) dated 6/2/22. Letter states reason for recall, health risk and action to take: These instructions apply only to clinics directly contacted by the Fresenius Medical Care Corrections and Removals Department. " Immediately examine your stock to determine whether you have NaturaLyte Liquid Acid concentrate with the part and lot numbers listed below. " Only bottles with sealed for your protection on the bottle cap foil are affected (see picture below). " Remove the affected product from your inventory. " FMCRTG, LLC will contact you regarding product return. Please contact FMCRTG Customer Service at 1-800-323-5188 if you have any further questions.
Recalling firm
- Firm
- Fresenius Medical Care Holdings, Inc.
- Address
- 920 Winter St, Bld 950, Waltham, Massachusetts 02451-1521
Distribution
- Distribution pattern
- Natiowide
Timeline
- Recall initiated
- 2022-05-31
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #194414. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.