FDA Device Recalls

Recalls /

#196277

Product

Cardinal Health Reusable Gel Pack, Hot/Cold Pack, 4.5 in. x 10 in., REF 70304A

FDA product code
IMEPack, Hot Or Cold, Reusable
Device class
Class 1
Medical specialty
Physical Medicine
Affected lot / code info
UDI/DI (01)30888499004898, All lots

Why it was recalled

Users are being informed that the product should not be used on infants/neonates. A caution statement will be added to the labeling stating: Not for use on Infants/Neonates .

Root cause (FDA determination)

Pending

Action the firm took

Cardinal Health sent an URGENT:MEDICAL DEVICE PRODUCT CORRECTION on 10/03/2022 by overnight mail. The notice explained the issue and requested the following: "Actions Required: 1. COMMUNICATE with all personnel that utilize these gel packs regarding appropriate use. 2. NOTIFY any customers to whom you may have distributed/forwarded affected product to or will send the product on to about this product correction notice and share a copy of this notice. 3. POST a copy of this notification in your storeroom where the product is stored 4. RETURN the enclosed acknowledgment form via fax to 614-652-9648 or email to gmbfieldcorrectiveaction@cardinalhealth.com, whether or not you have affected product." Additional Information: "For questions related to this notification and/or acknowledgement form that are not adequately addressed in this letter, please contact the market action team at: GMB-FieldCorrectiveAction@cardinalhealth.com or call 800-292-9332."

Recalling firm

Firm
Cardinal Health 200, LLC
Address
3651 Birchwood Dr, Waukegan, Illinois 60085-8337

Distribution

Distribution pattern
nationwide and Canada

Timeline

Recall initiated
2022-09-28
Posted by FDA
2022-11-07
Status