Product

Revaclear 300 dialyzers. Indicated for treatment of chronic and acute renal failure by hemodialysis.

FDA product code
KDIDialyzer, High Permeability With Or Without Sealed Dialysate System
Device class
Class 2
Medical specialty
Gastroenterology, Urology
510(k) numbers
K130039
Affected lot / code info
UDI-DI: 07332414123055; Product Code: 114745L; All lot numbers from C419128801-C422128401

Why it was recalled

Reports of Polyurethane Displacement in the dialyzer, and might be identified as a blood leak

Root cause (FDA determination)

Process control

Action the firm took

Baxter Healthcare notified healthcare facilities on 11/23/2022 and home patients on 11/29/2022, via first class mail. Consignees were instructed to be aware of the difference in appearances between Polyurethane Displacement and an Internal Blood Leak and ensure staff is also aware. Additionally, consignees were instructed that if there is a blood leak alarm, to follow your clinical practice to stop treatment, and if only the Polyurethane Displacement is observed, to continue use of the dialyzer. Consignees were asked to acknowledge receipt of the notification by completing and returning the response form.

Recalling firm

Firm
Baxter Healthcare Corporation
Address
1 Baxter Pkwy, Deerfield, Illinois 60015-4625

Distribution

Distribution pattern
Worldwide distribution - US Nationwide and the countries of Canada, China, Australia, New Zealand, Austria, Egypt, France, Germany, Ireland, Italy, Morocco, Qatar, Saudi Arabia, South Africa, Spain, United Kingdom, Bahamas, Barbados, Belize, Bermuda, Colombia, Dominican Republic, Ecuador, Guatemala, Honduras, Mexico, Panama, St. Thomas, Trinidad and Tobago.

Timeline

Recall initiated
2022-11-23
Posted by FDA
2022-12-22