Recalls / —
—#197597
Product
Luminos Agile Max, Luminos dRF Max, and LUMINOS Lotus Max
- FDA product code
- OWB — Interventional Fluoroscopic X-Ray System
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K173639
- Affected lot / code info
- All serial numbers 1) Luminos Agile Max, Material #10762472, UDI/DI: 4056869009162 2) Luminos drF Max, Material #10762471, UDI/DI: 4056869009155 3) Luminos Lotus Max, Material #11574100, UDI/DI: 4056869269331
Why it was recalled
There is potential for the footrest to detach from the patient table during use. Device intended to visualize anatomical structures by converting an Xray pattern into a visible image. The system has medical applications ranging from gastrointestinal examinations to cranial, skeletal, thoracic and lung exposures as well as examinations of the urogenital tract. The unit may also be used in emergency applications, lymphography, endoscopy, myelography, venography, arthrography, interventional radiology, digital angiography, and digital subtraction angiography (DSA). The system may be used on pediatric, adult, and bariatric patients. These devices are not for mammography examinations.
Root cause (FDA determination)
Process control
Action the firm took
A Customer Safety Advisory Notification (CSAN) was distributed to affected consignees beginning December 9, 2022. The CSAN was distributed to inform the customers about a preventative measure to exclude any potential risk of patient injury caused by the footrest detaching from the patient table during examination. Together with the CSAN, an addendum to the operator manual is distributed containing additional details and illustrations on how to securely attach the footrest to further improve the understanding of handling the footrest. Refer to the CSAN with addendum and copies of the Instructions for Use for each reported system.
Recalling firm
- Firm
- Siemens Medical Solutions USA, Inc
- Address
- 40 Liberty Blvd, Malvern, Pennsylvania 19355-1418
Distribution
- Distribution pattern
- Worldwide distribution
Timeline
- Recall initiated
- 2022-12-09
- Posted by FDA
- 2023-01-20
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #197597. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.