Recalls / —
—#198028
Product
Surgical convenience kits labeled as: CENTURION MEDICAL PRODUCTS IR INSTRUMENT TRAY, Reorder Number SUT10675
- FDA product code
- KYZ — Syringe, Irrigating (Non Dental)
- Device class
- Class 1
- Medical specialty
- General Hospital
- Affected lot / code info
- Reorder Number SUT10675, GTIN 40653160062845, Lot # 22HMG259
Why it was recalled
The recall is due to the perforated bag not allowing for outside of the kit wrapping to be sterile, therefore, only the contents inside the kits wrapping are sterile.
Root cause (FDA determination)
Under Investigation by firm
Action the firm took
The firm sent a correction notice to its direct consignees on 12/07/2022 by US mail and email. The notice explained the issue and requested the following: "Please check you stock for affected kits and over-labeling will be sent to place on kits to ensure that all labeling is accurate. No kits are to be returned back to Medline. All components are not affected. Our records indicate that these were used as a component in the list of kit(s) purchased from Medline Industries, LP." "If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Have the customers return any affected product to you for collection and return to Medline Industries, LP." "If you have any questions, please contact the Recall Department at 866-359-1704."
Recalling firm
- Firm
- MEDLINE INDUSTRIES, LP Northfield
- Address
- 3 Lakes Dr, Northfield, Illinois 60093-2753
Distribution
- Distribution pattern
- Nationwide
Timeline
- Recall initiated
- 2022-12-07
- Posted by FDA
- 2023-02-10
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #198028. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.