FDA Device Recalls

Recalls /

#198226

Product

Finished Goods Box of 50 Inserts and Scanning Labels for Medivac Guardian 2L Catalog Number: FG 12010. Intended adjunct in the estimation of blood loss.

FDA product code
PBZImage Processing Device For Estimation Of External Blood Loss
Device class
Class 2
Medical specialty
General Hospital
510(k) numbers
K142801
Affected lot / code info
UDI-DI: 10859506006125 Lot Numbers: M2110004, M2201002, M2202001, M2203002, M2206001

Why it was recalled

Triton Canister Software, Insert & Scanning Label may cause the Triton Canister System to output inaccurate blood loss estimates can lead to confusion and could potentially result in either delayed recognition or prolongation of postpartum hemorrhage (PPH), or require more aggressive treatment for PPH.

Root cause (FDA determination)

Under Investigation by firm

Action the firm took

Stryker issued Urgent Medical Device Correction letter to Risk Manager, OR Director, Materials Manager dated Jan 25, 2023. Letter states reason for recall, health risk and action to take: Steps: 1. Immediately check your internal inventory to locate any affected Finished Goods Inserts and Scanning labels (FG 12010 and FG 12022). Quarantine and discontinue use of all affected products. 2. A Stryker representative will contact your facility to assist updating the Triton Canister software. 3. Return the enclosed Business Reply Form (BRF) even if you don t have any affected product on hand, via email to instruments.recalls@stryker.com. 4. If the BRF indicates that recalled product is currently on hand, a shipping label will be provided to return recalled product. Upon receipt of the recalled product, Stryker will arrange replacements. 5. Maintain awareness of this communication internally until all required actions have been completed within your facility. Contact: Sylvia Page | Stryker Instruments Recall Coordinator 269-501-8375 |Instruments.recalls@stryker.com

Recalling firm

Firm
Stryker Corporation
Address
1941 Stryker Way, Portage, Michigan 49002-9711

Distribution

Distribution pattern
US Nationwide distribution.

Timeline

Recall initiated
2023-01-25
Posted by FDA
2023-02-27
Status

Source: openFDA Device Recall endpoint. Recall record ID #198226. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.