FDA Device Recalls

Recalls /

#199110

Product

SafeLight Fiber Optic Cable: Clear, 5MM X 10FT 3.05M, REF: 0233-050-100; Gray, 5MM X 10FT 3.05M, REF: 0233-050-200

FDA product code
FCWLight Source, Fiberoptic, Routine
Device class
Class 2
Medical specialty
Gastroenterology, Urology
510(k) numbers
K082813
Affected lot / code info
Ref/UDI-DI/Lot: 0233-050-100/07613327051650/159190, 159191, 159192, 159194, 159195, 159196, 159197, 159198, 159199, 159200, 159201, 159202, 159203, 159204, 159205, 159206, 159207, 159208, 159209, 159210, 159211, 159212, 159213, 159214, 159215, 159216, 159217, 159218, 164764, 164765, 164766, 164767, 164768, 164769, 164770, 164771, 164772, 164773, 164774, 164775, 164776, 164777, 164778, 164779, 164780, 164781, 164782, 164783, 164784, 164785, 164786, 164787, 164788, 164789, 164790, 164791, 164792, 164793, 164794, 164795, 164796, 164797, 164798, 164799, 164800, 164801, 164802, 164803, 164804, 164805, 164806, 164807, 164808, 164809, 164810, 164811, 164812, 164813, 164814, 164815, 164816, 164817, 164818, 164819, 164820, 164821, 164822, 168940, 168941, 168942, 168943, 168944, 168945, 168946, 168947, 168948, 168949, 168950, 168951, 168952, 168953, 168954, 168955, 168956, 168957, 168958, 168959, 168960, 168961, 168962, 168963, 168964, 168965, 168966, 168967, 168968, 168969, 168970, 168971, 168972, 168973, 168974, 168975, 168976, 168977, 168978, 168979, 168980, 168981, 168987; 0233-050-200/07613327051643/158865, 158868, 158869, 158872, 158874, 158875, 158878, 158884, 165446, 165448, 165451, 165457, 165458, 165459, 165461, 165462, 165464, 168909, 168910, 168911, 168912, 168913, 168914, 168915, 168916, 168917, 168918, 168919, 168920, 168921, 168923, 168924, 168925, 168926, 168927, 168928

Why it was recalled

Fiber optic cables assembled with not enough epoxy on the proximal end, creating light output issues, which could cause a clinical delay or procedure being converted to open.

Root cause (FDA determination)

Employee error

Action the firm took

On 3/2/23 recall notices were mailed to customers who were asked to do the following: 1. Inform individuals within your organization who need to be aware of this action. 2. Identify, remove, and quarantine affected devices. 3. If affected product is found, call customer service at 1-800-624-4422 (Option 3) or email endocustomersupport@stryker.com to arrange for product return. 4. Complete and return the Business Reply Form Customers with questions can contact EndoRecall@stryker.com

Recalling firm

Firm
Stryker Corporation
Address
5900 Optical Ct, San Jose, California 95138-1400

Distribution

Distribution pattern
Worldwide - US Nationwide distribution including in the states of PA, MS, IL, OH, CA, TX, MO, IN, KS, OK, MT, MA, NE, AZ, NV, NJ, IA, WV, MD, KY, FL, NM, NY, CO, WA, NC, VA, LA, ND, WY, HI SC, UT, OR, SD, AR, TN, AK, CT, GA, ID, VT, NH, MN, WI, DE, RI, AL, MI and the countries of NL, PE, AT, CA, TW, PH, SG, KR, CN.

Timeline

Recall initiated
2023-03-02
Posted by FDA
2023-04-28
Status