Recalls / —
—#199110
Product
SafeLight Fiber Optic Cable: Clear, 5MM X 10FT 3.05M, REF: 0233-050-100; Gray, 5MM X 10FT 3.05M, REF: 0233-050-200
- FDA product code
- FCW — Light Source, Fiberoptic, Routine
- Device class
- Class 2
- Medical specialty
- Gastroenterology, Urology
- 510(k) numbers
- K082813
- Affected lot / code info
- Ref/UDI-DI/Lot: 0233-050-100/07613327051650/159190, 159191, 159192, 159194, 159195, 159196, 159197, 159198, 159199, 159200, 159201, 159202, 159203, 159204, 159205, 159206, 159207, 159208, 159209, 159210, 159211, 159212, 159213, 159214, 159215, 159216, 159217, 159218, 164764, 164765, 164766, 164767, 164768, 164769, 164770, 164771, 164772, 164773, 164774, 164775, 164776, 164777, 164778, 164779, 164780, 164781, 164782, 164783, 164784, 164785, 164786, 164787, 164788, 164789, 164790, 164791, 164792, 164793, 164794, 164795, 164796, 164797, 164798, 164799, 164800, 164801, 164802, 164803, 164804, 164805, 164806, 164807, 164808, 164809, 164810, 164811, 164812, 164813, 164814, 164815, 164816, 164817, 164818, 164819, 164820, 164821, 164822, 168940, 168941, 168942, 168943, 168944, 168945, 168946, 168947, 168948, 168949, 168950, 168951, 168952, 168953, 168954, 168955, 168956, 168957, 168958, 168959, 168960, 168961, 168962, 168963, 168964, 168965, 168966, 168967, 168968, 168969, 168970, 168971, 168972, 168973, 168974, 168975, 168976, 168977, 168978, 168979, 168980, 168981, 168987; 0233-050-200/07613327051643/158865, 158868, 158869, 158872, 158874, 158875, 158878, 158884, 165446, 165448, 165451, 165457, 165458, 165459, 165461, 165462, 165464, 168909, 168910, 168911, 168912, 168913, 168914, 168915, 168916, 168917, 168918, 168919, 168920, 168921, 168923, 168924, 168925, 168926, 168927, 168928
Why it was recalled
Fiber optic cables assembled with not enough epoxy on the proximal end, creating light output issues, which could cause a clinical delay or procedure being converted to open.
Root cause (FDA determination)
Employee error
Action the firm took
On 3/2/23 recall notices were mailed to customers who were asked to do the following: 1. Inform individuals within your organization who need to be aware of this action. 2. Identify, remove, and quarantine affected devices. 3. If affected product is found, call customer service at 1-800-624-4422 (Option 3) or email endocustomersupport@stryker.com to arrange for product return. 4. Complete and return the Business Reply Form Customers with questions can contact EndoRecall@stryker.com
Recalling firm
- Firm
- Stryker Corporation
- Address
- 5900 Optical Ct, San Jose, California 95138-1400
Distribution
- Distribution pattern
- Worldwide - US Nationwide distribution including in the states of PA, MS, IL, OH, CA, TX, MO, IN, KS, OK, MT, MA, NE, AZ, NV, NJ, IA, WV, MD, KY, FL, NM, NY, CO, WA, NC, VA, LA, ND, WY, HI SC, UT, OR, SD, AR, TN, AK, CT, GA, ID, VT, NH, MN, WI, DE, RI, AL, MI and the countries of NL, PE, AT, CA, TW, PH, SG, KR, CN.
Timeline
- Recall initiated
- 2023-03-02
- Posted by FDA
- 2023-04-28
- Status
- —