FDA Device Recalls

Recalls /

#199600

Product

stryker ORTHOLOC 3Di Fusion Plate MTP Right, REF 58A220RT, bone stabilization and fixation

FDA product code
HRSPlate, Fixation, Bone
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K120359
Affected lot / code info
UDI/DI 00889797101011, Lot Number 1736809

Why it was recalled

Two lots of Ortholoc 3Di Fusion plates contain incorrect dorsiflexion descriptions laser marked on the parts.

Root cause (FDA determination)

Employee error

Action the firm took

The firm issued an URGENT: MEDICAL DEVICE RECALL notice to its sales representatives and distributors on 04/10/2023 by email and followed with letters disseminated to the remaining consignees via tracked FedEx mailing on Wednesday, April 12. The notice explained the issue and the risk and requested the following actions be taken: Identify all affected products and isolate/quarantine to prevent accidental use. All parties that further distributed the product were directed to notify anyone to the recipients of the affected products. The firm is seeking return of the products.

Recalling firm

Firm
Wright Medical Technology, Inc.
Address
1023 Cherry Rd, Memphis, Tennessee 38117-5423

Distribution

Distribution pattern
US Nationwide distribution.

Timeline

Recall initiated
2023-04-10
Posted by FDA
2023-04-26
Status

Source: openFDA Device Recall endpoint. Recall record ID #199600. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.