FDA Device Recalls

Recalls /

#199867

Product

Incisive CT for Brazil SKD-Computed Tomography X-Ray System Model: 728146

FDA product code
JAKSystem, X-Ray, Tomography, Computed
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K180015
Affected lot / code info
UDI-DI: (01)00884838104481(21) Serial Numbers: 540009 540008 540015 540012 540016 540017 540003 540004 540005 540006 540007 540001 540011 540000 540014 540013 540018 540010 540002

Why it was recalled

Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down

Root cause (FDA determination)

Under Investigation by firm

Action the firm took

Phiips issued Urgent Medical Device Correction Letters mailed to US consignees via Certified United States Postal Service (USPS) with delivery confirmation on 4/25/23 International global markets were notified of the issue and instructed to distribute the FSN to affected consignees in accordance with local requirements. Letter states reason for recall, health risk and action to take: Please continue to use your system in accordance with its intended use. If you need to remain in the room for the duration of a scan, Philips recommends using protective eyewear. " Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. Please retain this letter with your system(s) until a solution is installed on your system; ensure the letter is in a place likely to be seen/viewed. " Place this URGENT Medical Device Correction Letter with your system documentation. " Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken. Philips will contact you to schedule a time for a Field Service Engineer (FSE) to visit your site and install a solution to address the issue (reference FCO72800798). If you need additional information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).

Recalling firm

Firm
Philips North America
Address
222 Jacobs St, Cambridge, Massachusetts 02141-2296

Distribution

Distribution pattern
Nationwide Foreign: China Algeria Argentina Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Canada Chile China Colombia Czech Republic Ecuador Egypt France Germany Greece India Indonesia Iran Iraq Israel Italy Japan Jordan Korea, Republic of Kosovo Latvia Macedonia Mexico Moldova Netherlands Nigeria Pakistan Panama Philippines Poland Romania Russian Federation Saint Lucia Saint Pierre and Miquelon Saudi Arabia Serbia Slovakia Spain Sweden Taiwan Thailand Tunisia Turkey Turkmenistan United Kingdom Uzbekistan Viet Nam

Timeline

Recall initiated
2023-04-25
Posted by FDA
2023-07-28
Status

Source: openFDA Device Recall endpoint. Recall record ID #199867. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.