Product
Snowden-Pencer MIS DIAMOND-TOUCH CLAMP DEBAKEY 25 MM JAW 5MM, 45CM, RATCHETED, REF SP90-7243; endoscopic surgical device
- FDA product code
- GCJ — Laparoscope, General & Plastic Surgery
- Device class
- Class 2
- Medical specialty
- Gastroenterology, Urology
- 510(k) numbers
- K930667
- Affected lot / code info
- UDI/DI 10885403161384, Lot Number B23
Why it was recalled
Specific lots of the product have the potential to exhibit cracking or breaking of the rotation knob due to the incorrect use of an adhesive in the manufacturing process.
Root cause (FDA determination)
Nonconforming Material/Component
Action the firm took
BD issued an URGENT: Medical Device Product Recall notice to its consignees on 13 July 2023 via FedEx 2-day and email. The notice explained the problem with the device, the risk to the patient, and requested the following actions: Medical facilities: - Discontinue the use of the affected device and destroy it - Share the notice with all users within your facility network - If the product was purchased through a distributor, contact the distributor directly for further instructions. Distributors were directed to Identify all customers within your distribution network that purchased any affected product as defined in this notification. Provide a copy of the attached customer letter to all customers to advise them of this recall notification on BD s behalf. Questions: Phone: 1-844-8BD- LIFE (1-844-823-5433), Say Recall when prompted, M-F 8am - 5pm CT Email: productcomplaints@bd.com
Recalling firm
- Firm
- Carefusion 2200 Inc 75 N Fairway Dr Vernon Hills IL 60061-1845
Distribution
- Distribution pattern
- US and Canada
Timeline
- Recall initiated
- 2023-07-13
- Posted by FDA
- 2024-09-30
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #201702. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.