Recalls / —
—#201743
Product
DYND3000xxP Series - Flexible Laryngeal Mask Airway Packs
- FDA product code
- CAE — Airway, Oropharyngeal, Anesthesiology
- Device class
- Class 1
- Medical specialty
- Anesthesiology
- Affected lot / code info
- All Lots for series since product launch; Model Number (Unique Device Identifier) -DYND300010P (Each level packaging: 10193489010671; Case level packaging: 20193489010678); DYND300015P (Each level packaging: 10193489010688; Case level packaging: 20193489010685); DYND300020P (Each level packaging: 10193489010695; Case level packaging: 20193489010692); DYND300025P (Each level packaging: 10193489010701 Case level packaging: 20193489010708); DYND300030P (Each level packaging: 10193489010718; Case level packaging: 20193489010715); DYND300040P (Each level packaging: 10193489010725; Case level packaging: 20193489010722); DYND300050P (Each level packaging: 10193489010732; Case level packaging: 20193489010739)
Why it was recalled
The case and each (i.e. packet) labels for the DYND3000xxP series do not include an expiration date, however, the lubricating jelly (MDS032280) included in the DYNDP3000xxP series is labeled with an expiration date.
Root cause (FDA determination)
Labeling Change Control
Action the firm took
The consignee recall notification was sent out via first class mail and email on 03/28/2023. The letter instructs the consignee to over-label the Flexible Laryngeal Mask Airway Packs stating to Check the Expiration date of Lubricating Jelly MDS032280 and do not use if Expired . No product is to be returned back to Medline Industries, LP.
Recalling firm
- Firm
- MEDLINE INDUSTRIES, LP Northfield
- Address
- 3 Lakes Dr, Northfield, Illinois 60093-2753
Distribution
- Distribution pattern
- Worldwide distribution - US Nationwide and the country of Bermuda.
Timeline
- Recall initiated
- 2023-03-28
- Posted by FDA
- 2023-07-21
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #201743. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.