Product

Philips Allura Xper, Allura Centron, Allura CV, Azurion, UNIQ Clarity, and MultiDiagnost-Eleva systems

FDA product code
OWBInterventional Fluoroscopic X-Ray System
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K050151, K162859, K163715, K200917
Affected lot / code info
Models: 708032, 708034, 708036, 708037, 708038, 722001, 722002, 722003, 722005, 722006, 722008, 722010, 722011, 722012, 722013, 722015, 722019, 722020, 722022, 722023, 722025, 722026, 722027, 722028, 722029, 722031, 722033, 722035, 722038, 722039, 722058, 722059, 722063, 722064, 722065, 722066, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228, 722280, 722400 No UDI-DI information for foot switch component.

Why it was recalled

A damaged foot switch may cause instances where no or only intermittent x-ray radiation initiation is possible.

Root cause (FDA determination)

Device Design

Action the firm took

US customers were notified via certified mail via USPS. OUS customers were notified through the Philips Markets Organizations. Philips is issuing two Urgent Medical Device Correction letters, one for Allura and Azurion customers and another one for MultiDiagnost Eleva Customers. Additionally, an IFU addendum will be provided to customers that includes instructions on how to handle the footswitch cable. For Allura and Azurion customers an updated preventative maintenance manual update is also being provided. Philips will contact customers to schedule a visit to inspect the foot switch and to provide an IFU Addendum.

Recalling firm

Firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis 4-6 Best Netherlands

Distribution

Distribution pattern
Worldwide distribution

Timeline

Recall initiated
2023-08-15
Posted by FDA
2023-09-14
Status

Source: openFDA Device Recall endpoint. Recall record ID #202848. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.