FDA Device Recalls

Recalls /

#203335

Product

BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Numbers BF-1T150, BF-1T60, BF-P150 BF-P60 and EVIS EXERA BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Number XT160

FDA product code
EOQBronchoscope (Flexible Or Rigid)
Device class
Class 2
Medical specialty
Ear, Nose, Throat
510(k) numbers
K023984
Affected lot / code info
Model BF-1T150 UDI-DI: 4953170308185; Model BF-1T60 UDI-DI: 4953170339264; Model BF-P150 UDI-DI: N/A; Model BF-P60 UDI-DI: 4953170339196; Model BF-XT160 UDI-DI: N/A

Why it was recalled

There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.

Root cause (FDA determination)

Under Investigation by firm

Action the firm took

An "URGENT MEDICAL DEVICE CORRECTIVE ACTION" notification letter dated 10/12/23 was sent to customers. Actions to be taken by the end user: Our records indicate that your facility has purchased one or more of the affected bronchoscopes. Olympus requests you to take the following actions: 1. Inspect your facility for the referenced devices and ensure all personnel are completely knowledgeable and thoroughly aware of the Warnings in affected bronchoscope s Operation Manual for use with High-frequency therapy equipment and that Olympus high-frequency compatible bronchoscopes are compatible only with Combination equipment list in operation manual. You may continue to use the device according to the existing instructions and warnings contained in the Instruction for Use. 2. Complete the enclosed response form and return to our recall partner, Sedgwick, via Email (Olympus4907@sedgwick.com) or Fax (866-808-1177). For any questions about the acknowledgement form, please call the Sedgwick team at 855-215-5128. 3. If you have further distributed this product outside of your facility, forward them this notification and appropriately document your notification process. Olympus requests that you report to Olympus complaints, including any injuries associated with procedures involving energy devices used with Olympus bronchoscopes. Please report complaints to the Olympus Technical Assistance Center (TAC) at 1-800-848-9024 (option 1). Adverse events experienced with the use of this product may also be reported to the Food and Drug Administration MedWatch Adverse Event Reporting program either online, by regular mail or by fax. Olympus regrets any inconvenience caused and fully appreciates your cooperation in this matter. Please do not hesitate to contact me directly at (647) 999-3203 or at Cynthia.Ow@Olympus.com for any additional information or support concerning this matter.

Recalling firm

Firm
Olympus Corporation of the Americas
Address
3500 Corporate Pkwy, Center Valley, Pennsylvania 18034-8229

Distribution

Distribution pattern
US Nationwide distribution

Timeline

Recall initiated
2023-10-12
Posted by FDA
2023-11-09
Status

Source: openFDA Device Recall endpoint. Recall record ID #203335. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.