FDA Device Recalls

Recalls /

#205129

Product

Custom procedural convenience kits and trays, gastroenterological & urological, labeled as: a) LAP KIDNEY MODULE, kit number AGKD50L; b) ROBOTIC PROSTATE PACK - 264852, kit number ANPR34S; c) LAP CHOLE - 213388, kit number BFLC02AD; d) GENERAL LAPAROSCOPY PACK, kit number GUGE09I; e) LAP CHOLE PACK, kit number HNLC80I; f) Lap Chole Ovarian Appy, kit number JSLC06I; g) LAP CHOLE, ST LUKES, kit number LMLP08Z; h) ROBOTIC PACK - 242561, kit number MHRB88AJ; i) KIT, GENERAL LAPAROSCOPY, kit number MMLP40F; j) KIT, MAJOR LAPAROTOMY, kit number MMML26F; k) LAP CHOLE - LAP HERNIA PACK - 282807, kit number OWLC07K; l) GENERAL LAPAROSCOPY SAH, kit number SAGL57H; m) ROBOTIC PACK SAH, kit number SARB30K; n) ROBOTIC PACK SAH, kit number SARB30L; o) ROBOTIC PACK SAH, kit number SARB30M; p) LAP CHOLE PACK, kit number SMLC53; q) ROBOTICS PACK, kit number SNRP85D; r) ROBOTIC PACK, kit number SSRO22C; s) ROBOTIC PACK, kit number SSRO22D; t) TPK DAVINCI PROSTATE SYSTEM, kit number TPDV16; u) ASC LAP CHOLE PACK (PS 42279), kit number UICH53Y; v) GENERAL LAPAROSCOPIC PACK, kit number WVGL60E

FDA product code
FDELaparoscopy Kit
Device class
Class 2
Medical specialty
Gastroenterology, Urology
Affected lot / code info
a) AGKD50L, UDI/DI 00191072141156, Lot Numbers: 905231, exp. 4/5/2024; 638221, exp. 12/28/2023; 689221, exp. 5/21/2023; b) ANPR34S, UDI/DI 00191072171351, Lot Numbers: 721221, exp. 10/6/2023; 879231, exp. 5/1/2024; 896231, exp. 4/14/2024; 927231, exp. 3/14/2024; 956231, exp. 2/13/2024; 645221, exp. 12/21/2023; c) BFLC02AD, UDI/DI 00191072169075, Lot Numbers: 654221, exp. 12/12/2023; 647221, exp. 12/19/2023; 987231, exp. 1/13/2024; UDI/DI 00191072188830, Lot Numbers: 946231, exp. 2/23/2024; 896231, exp. 4/14/2024; d) GUGE09I, UDI/DI none, Lot Numbers: 768221, exp. 8/20/2023; e) HNLC80I, UDI/DI 00191072146267, Lot Numbers: 797221, exp. 7/22/2023; 679221, exp. 11/17/2023; 995231, exp. 1/25/2024; 911231, exp. 3/30/2024; f) JSLC06I, UDI/DI 00191072126061, Lot Numbers: 839221, exp. 6/10/2023; 766221, exp. 8/22/2023; g) LMLP08Z, UDI/DI none, Lot Numbers: 661221, exp. 12/5/2023; 987231, exp. 1/13/2024; 942231, exp. 2/27/2024; 893231, exp. 4/17/2024; h) MHRB88AJ, UDI/DI none, Lot Numbers: 735221, exp. 9/22/2023; 710221, exp. 10/17/2023; 997231, exp. 1/3/2024; 959231, exp. 2/10/2024; 904231, exp. 4/6/2024; i) MMLP40F, UDI/DI none, Lot Numbers: 933231, exp. 3/8/2024; 879231, exp. 5/1/2024; j) MMML26F, UDI/DI none, Lot Numbers: 989231, exp. 1/11/2024; 896231, exp. 4/14/2024; k) OWLC07K, UDI/DI none, Lot Numbers: 682221, exp. 11/14/2023; 646221, exp. 12/20/2023; 982231, exp. 1/18/2024; 928231, exp. 3/13/2024; l) SAGL57H, UDI/DI none, Lot Numbers: 687221, exp. 11/9/2023; 790221, exp. 7/29/2023; m) SARB30K, UDI/DI none, Lot Numbers: 845221, exp. 6/4/2023; 888221, exp. 4/22/2023; n) SARB30L, UDI/DI none, Lot Numbers: 752221, exp. 9/5/2023; o) SARB30M, UDI/DI none, Lot Numbers: 658221, exp. 12/8/2023; 700221, exp. 10/27/2023; 792221, exp. 7/27/2023; 949231, exp. 2/20/2024; 983231, exp. 1/17/2024; p) SMLC53, UDI/DI 00191072188465, Lot Numbers: 969231, exp. 1/31/2024; q) SNRP85D, UDI/DI none, Lot Numbers: 666221, exp. 11/30/2023; 720221, exp. 10/7/2023; 793221, exp. 7/26/2023; 853221, exp. 5/27/2023; 965231, exp. 2/4/2024; r) SSRO22C, UDI/DI none, Lot Numbers: 891221, exp. 04/19/2023; s) SSRO22D, UDI/DI none, Lot Numbers: 660221, exp. 5/31/2023; 698221, exp. 10/29/2023; 756221, exp. 9/1/2023; 804221, exp. 7/15/2023; 840221, exp. 6/9/2023; 933231, exp. 8/31/2023; 987231, exp. 1/13/2024; t) TPDV16, UDI/DI 00191072128904, Lot Numbers: 856221, exp. 5/24/2023; 744221, exp. 9/13/2023; 675221, exp. 11/21/2023; 932231, exp. 3/9/2024; u) UICH53Y, UDI/DI none, Lot Numbers: 897231, exp. 4/13/2024; 882231, exp. 4/28/2024; v) WVGL60E, UDI/DI 00191072144034, Lot Numbers: 766221, exp. 8/22/2023; 694221, exp. 11/2/2023; 952231, exp. 2/17/2024;

Why it was recalled

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

Root cause (FDA determination)

Process control

Action the firm took

Owens & Minor issued an URGENT: MEDICAL DEVICE CORRECTION notice to it consignees on 11/29/2023 via email. Then notice explained the problem, risk to health, and requested the following actions be taken: If you received any of the batches/lots identified, please segregate and quarantine all affected product. Add warning labels to all kits/cases impacted: A labeling template is enclosed for printing warning labels which should be affixed to all affected product in inventory. This label should be affixed to each affected kit and case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. " Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. " Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified of this field correction and maintain records of effectiveness. Report any quality problems or adverse events associated with the products listed in this notification to complaints@owens-minor.com.

Recalling firm

Firm
American Contract Systems, Inc.
Address
7802 E Telecom Pkwy, Temple Terrace, Florida 33637-0928

Distribution

Distribution pattern
US Nationwide distribution.

Timeline

Recall initiated
2023-11-29
Posted by FDA
2024-01-26
Status