Recalls / —
—#205535
Product
BD BBL Sensi Disc Minocycline - 30 ug, used for semi-quantitative in vitro susceptibility testing. Catalog Number 231250 and 231251
- FDA product code
- JTN — Susceptibility Test Discs, Antimicrobial
- Device class
- Class 2
- Medical specialty
- Microbiology
- 510(k) numbers
- K802194
- Affected lot / code info
- Catalog No. 231250 UDI-DI 00382902312505 Lots 1085656 1277053 1301694 1334451 2025108 2119580 2146396 3181025; Catalog No. 231251¿ UDI-DI 30382902312513 Lots 1007140 2146396 3181025 3234149 1085656 1210753 1277053 1301694 1334451 2025108 2119580
Why it was recalled
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
Root cause (FDA determination)
Process control
Action the firm took
On January 8, 2024, the firm notified customers through URGENT Medical Device Product Correction letters. Users were instructed to cease use of catalog number 291270, and to refrain from conducting AST for H. influenzae with impacted Sensi-Discs. Customers should destroy any units of catalog number 291270 in their inventory. Customers should share and post the recall notification within their facility network and forward to any customers to whom they may have distributed the product. Future lots of affected Sensi-Disc products (excluding catalog number 291270) will be labeled with a sticker notifying customers "This product should not be used for the semi-quantitative in vitro susceptibility testing of Haemophilus influenzae." Identify impacted lot numbers: https://legacy.bd.com/alerts-notices/IDS-23-4851.asp For customer questions: Phone: 1-844-823-5433 Say "Recall when prompted Mon-Fri 8:00am and 5:00pm CT or Email: productcomplaints@bd.com
Recalling firm
- Firm
- Becton Dickinson & Co.
- Address
- 7 Loveton Cir, Sparks, Maryland 21152-9212
Distribution
- Distribution pattern
- Domestic distribution nationwide. International distribution worldwide.
Timeline
- Recall initiated
- 2024-01-08
- Posted by FDA
- 2024-02-08
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #205535. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.