Recalls / —
—#206452
Product
Centurion PICC LINE SECUREMENT KIT, Product Code IVS565
- FDA product code
- OXQ — Dressing Change Tray
- Device class
- Class 1
- Medical specialty
- General Hospital
- Affected lot / code info
- UDI/DI 10653160175971 (case) 00653160175974 (each), Lot Numbers: 2023070790
Why it was recalled
Product may have a weak seal which may result in a breach of sterility of the contents if the seal fails. The weak seal may not be detectable by all users.
Root cause (FDA determination)
Process control
Action the firm took
Medline Industries, LP, issued a Medical Device Recall notice to its consignees on 02/27/2024 via email and USPS First Class mail. The notice explained the problem with the devices, risk, and requested that the products be destroyed at the consignee. Distributors were directed to notify their customers.
Recalling firm
- Firm
- MEDLINE INDUSTRIES, LP Northfield
- Address
- 3 Lakes Dr, Northfield, Illinois 60093-2753
Distribution
- Distribution pattern
- Worldwide distribution - US Nationwide and the countries of Panama, Canada.
Timeline
- Recall initiated
- 2024-02-23
- Posted by FDA
- 2024-03-29
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #206452. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.