FDA Device Recalls

Recalls /

#206798

Product

Medtronic Aortic Root Cannulae and Cardioplegia Needles: a) DLP 9 Ga (11 Fr) Aortic Root Cannula, Model Number 24009; b) DLP 16 Ga (5 Fr) Cardioplegia Needle Neonatal 0.64 cm (1/4 in) Tip Length, Model Number 11316; c) M+AR" 14 Ga (7 Fr) Aortic Root Cannula with Flow-Guard", Model Number 11014L

FDA product code
DWFCatheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Device class
Class 2
Medical specialty
Cardiovascular
510(k) numbers
K182586
Affected lot / code info
a) Model Number 24009, GTIN 00673978178646, Lot Numbers: 2023040211, 2023040854, 2023041224; a) Model Number 24009, GTIN 20613994495420, Lot Numbers: 2023031043, 2023031563, 2023040211, 2023040480, 2023040854, 2023040855, 2023041224, 202305C108, 202305C109, 202306C196; b) Model Number 11316, GTIN 20613994676140, Lot Numbers: 2023040437; c) Model Number 11014L, GTIN 20613994495499, Lot Numbers: 2023040807, 2023040808, 202305C076, 202306C138

Why it was recalled

Potential for unsealed sterile packing.

Root cause (FDA determination)

Process control

Action the firm took

Medtronic issued an Urgent Medical Device Recall notice to its consignees on 03/19/2024 via letter (UPS). The notice explained the issue, potential risk, and requested the consignee identify and return the product. Consignee was directed to forward the notice to all affected parties. For questions, the consignee was directed to contact their Medtronic field representative.

Recalling firm

Firm
Medtronic Perfusion Systems
Address
7611 Northland Dr N, Brooklyn Park, Minnesota 55428-1088

Distribution

Distribution pattern
Worldwide distribution.

Timeline

Recall initiated
2024-03-19
Posted by FDA
2024-05-10
Status

Source: openFDA Device Recall endpoint. Recall record ID #206798. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.