Recalls / —
—#208575
Product
Medline Convenience kits, labeled as: 1) TRACHEOTOMY TRAY, Pack Number DYNDA1894; 2) ULTRASOUND GUIDED PIV KIT, Pack Number MNS9580;
- FDA product code
- BSY — Catheters, Suction, Tracheobronchial
- Device class
- Class 1
- Medical specialty
- Anesthesiology
- Affected lot / code info
- DYNDA1894, Lot Number 2019080890; DYNDA1894, Lot Number 2019082190; MNS9580, Lot Number 2019080890
Why it was recalled
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Root cause (FDA determination)
Under Investigation by firm
Action the firm took
Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.
Recalling firm
- Firm
- MEDLINE INDUSTRIES, LP Northfield
- Address
- 3 Lakes Dr, Northfield, Illinois 60093-2753
Distribution
- Distribution pattern
- Worldwide distribution.
Timeline
- Recall initiated
- 2024-04-08
- Posted by FDA
- 2024-09-06
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #208575. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.