Product

Medline Convenience kits labeled as: 1) DIALYSIS CATHETER INSERTION KIT, Pack Number CVI4815; 2) ANGIO PACK, Pack Number CVI5045; 3) CHILDRENS INFANT SP TRAY, Pack Number DYNDA2906; 4) ULTRASOUND GUIDE BREAST BIOPSY, Pack Number DYNDH1319; 5) BIOPSY TRAY, Pack Number DYNDH1693A; 6) PACK,ULTRASOUND, Pack Number DYNDL1461; 7) LUMBAR PUNCTURE TRAY, Pack Number DYNJTS4306; 8) ARTHROGRAM TRAY, Pack Number SPEC0143A; 9) PD CATH KIT, Pack Number UVT1205

FDA product code
OFDCatheter Introducer Kit
Device class
Class 2
Medical specialty
Cardiovascular
Affected lot / code info
CVI4815, Lot Number 2022031150; CVI4815, Lot Number 2022042250; CVI4815, Lot Number 2022121390; CVI5045, Lot Number 2023101290; DYNDA2906, Lot Number 2022042250; DYNDH1319, Lot Number 2022031150; DYNDH1693A, Lot Number 2022042250; DYNDL1461, Lot Number 2022042250; DYNJTS4306, Lot Number 2022042250; SPEC0143A, Lot Number 2022042250; UVT1205, Lot Number 2022072650; UVT1205, Lot Number 2022081650

Why it was recalled

Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.

Root cause (FDA determination)

Under Investigation by firm

Action the firm took

Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.

Recalling firm

Firm
MEDLINE INDUSTRIES, LP Northfield
Address
3 Lakes Dr, Northfield, Illinois 60093-2753

Distribution

Distribution pattern
Worldwide distribution.

Timeline

Recall initiated
2024-04-08
Posted by FDA
2024-09-06
Status