Product

Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Non-Wireless Model/Catalog Number: 8713032U Product Description: Perfusor Space Syringe Pump, Non-Wireless, Infusion Pump, 1 each Component: N/A

FDA product code
FRNPump, Infusion
Device class
Class 2
Medical specialty
General Hospital
510(k) numbers
K172831, K191910
Affected lot / code info
Lot Code: Model No 8713032U; UDI-DI 04046964797965; All Serial Numbers

Why it was recalled

Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.

Root cause (FDA determination)

Device Design

Action the firm took

On June 14, 2024, the firm notified customers via URGENT MEDICAL DEVICE FIELD CORRECTION NOTIFICATION letters. Customers were advised to not use the recalled syringes in Perfusor Space Syringe Pumps. All users should be informed of the correction. Customers were provided with an attachment to post in care areas as a reminder to not use the recalled syringes with the specified pumps.

Recalling firm

Firm
B Braun Medical Inc
Address
200 Boulder Dr, Breinigsville, Pennsylvania 18031-1532

Distribution

Distribution pattern
Domestic distribution nationwide. International distribution to Canada.

Timeline

Recall initiated
2024-06-14
Posted by FDA
2024-07-25