Product

Medline Convenience kits labeled as: 1) SH STRL UMBILICAL ACC PACK-LF, Pack Number DYNJ0621408V; 2) SH STRL UMBILICAL ACC PACK-LF, Pack Number DYNJ0621408VH; 3) 6" ACE BANDAGE PACK-LF, Pack Number DYNJ0864492A ; 4) 6" ACE BANDAGE PACK-LF, Pack Number DYNJ0864492AH; 5) KIT ANGIO CSTM-PROVIDENCE , Pack Number SAMP0545

FDA product code
FQMBandage, Elastic
Device class
Class 1
Medical specialty
General Hospital
Affected lot / code info
DYNJ0621408V, Lot Number 23ILB053; DYNJ0621408VH, Lot Number 23ILB053; DYNJ0864492A , Lot Number 21KMH585 ; DYNJ0864492AH, Lot Number 21KMH585 ; SAMP0545 , Lot Number 22GME387 ; SAMP0545 , Lot Number 22GME980

Why it was recalled

Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.

Root cause (FDA determination)

Under Investigation by firm

Action the firm took

Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.

Recalling firm

Firm
MEDLINE INDUSTRIES, LP Northfield
Address
3 Lakes Dr, Northfield, Illinois 60093-2753

Distribution

Distribution pattern
Worldwide distribution.

Timeline

Recall initiated
2024-04-08
Posted by FDA
2024-09-06
Status