Recalls / —
—#208877
Product
BD PYXIS MEDBANK TWR MN CR-12HH-1FH, REF: 169-84 containing software version 3.9.1.9
- FDA product code
- BRY — Cabinet, Table And Tray, Anesthesia
- Device class
- Class 1
- Medical specialty
- Anesthesiology
- Affected lot / code info
- UDI: 10885403512568/ Serial Number: 16167616 16188929 15941657 15941700 15941694 15922822 15941698
Why it was recalled
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
Root cause (FDA determination)
Under Investigation by firm
Action the firm took
On July 11, 2024, CareFusion, subsidiary of Becton, Dickinson and Company (BD) issued a "Urgent Medical Device Correction" Notification, via mail. On June 24, 2024, BD issued Customer Release Notes via their firm's customer portal. BD ask consignees to take the following actions: Clinicians: 1. When restocking the cabinet, refer to the MedBank restock workflow screen to verify the correct bin location rather than the printed label until the software update is provided. Users should verify the contents of the bin location being restocked. 2. Please circulate this notice to all those who require awareness within your organization or to any organization whom you may have transferred BD Pyxis" MedBank" CUBIE" Replenishment Station devices to. 3. Please promptly complete and return the enclosed Customer Response Form within 15 days to acknowledge receipt of this notification and that you understand the information provided in this letter Distributor: 1. Please provide a copy of the attached customer letter to all facilities within your distribution network that have any product listed in Appendix 1. 2. Please promptly complete and return the enclosed Distributor Response Form within 15 days to acknowledge receipt of this notification and that you understand the information provided in this letter.
Recalling firm
- Firm
- CareFusion 303, Inc.
- Address
- 10020 Pacific Mesa Blvd, San Diego, California 92121-4386
Distribution
- Distribution pattern
- US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV.
Timeline
- Recall initiated
- 2024-06-24
- Posted by FDA
- 2024-08-02
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #208877. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.