FDA Device Recalls

Recalls /

#209520

Product

ANESTHESIA TURNOVER KIT, Pack Number DYNJAA0377A

FDA product code
OJHOrthopedic Tray
Device class
Class 1
Medical specialty
General, Plastic Surgery
Affected lot / code info
UDI/DI Case 40195327334209, UDI/DI Each 10195327334208; Lot Numbers: 23IBJ309, 23GBA500, 23GBA457, 23CBE394, 23BBL312

Why it was recalled

A non-sterile Mastisol component, which was not labeled as being sterile or non-sterile, was placed in the non-sterile portion of certain sterile convenience kits and distributed. In the event that the user assumes sterility during its use, there is an increased risk of patient infection.

Root cause (FDA determination)

Process change control

Action the firm took

Medline Industries issued a Recall notice to its consignees on 06/18/2024 via email. The notice explained the issue, potential risk, and directed the product to be stickered to instruct the used to remove the recalled component from the kit and replace from sterile supply. Distributors or those who have transferred the product were directed to notify those to whom the product was distributed of the recall. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.

Recalling firm

Firm
MEDLINE INDUSTRIES, LP Northfield
Address
3 Lakes Dr, Northfield, Illinois 60093-2753

Distribution

Distribution pattern
US: TX, IL, CA

Timeline

Recall initiated
2024-06-18
Posted by FDA
2024-09-06
Status

Source: openFDA Device Recall endpoint. Recall record ID #209520. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.