FDA Device Recalls

Recalls /

#209527

Product

Medline convenience kits containing semi-rigid suction liners labeled as follows: a) ENDO BRONCH PACK, Pack Number DYKE1355; b) ENDO BASIC KIT, Pack Number DYKE1356; c) ENDOSCOPY KIT, Pack Number DYKE1771C

FDA product code
NWUEndoscope Introducer Kit
Device class
Class 2
Medical specialty
Gastroenterology, Urology
Affected lot / code info
a) DYKE1355: UDI/DI case 40889942274193, UDI/DI each 10889942274192; Lot Numbers: 23LBB043, 24BBJ327, 24CBS751; b) DYKE1356: UDI/DI case 40889942274186, UDI/DI each 10889942274185; Lot Numbers: 23LBD062, 24ABA646, 24BBJ328, 24CBG449, 24DBE106, 24EBG238; c) DYKE1771C: UDI/DI case 40195327331994, UDI/DI each 10195327331993; Lot Numbers: 24ABX292

Why it was recalled

Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.

Root cause (FDA determination)

Process control

Action the firm took

Medline Industries issued a MEDICAL DEVICE RECALL notice to its consignees on 06/27/2024 via US mail and email. The notice explained the problem with the device, potential risk to the patient, and requested the consignee immediately check their stock for the affected item number and the affected lot numbers and destroy any affected inventory. Distributors were directed to notify their customers of the recall and request the affected products be destroyed. For questions: contact the Recall Department at 866-359-1704 or recalls@medline.com.

Recalling firm

Firm
MEDLINE INDUSTRIES, LP Northfield
Address
3 Lakes Dr, Northfield, Illinois 60093-2753

Distribution

Distribution pattern
Worldwide distribution - US Nationwide and the countries of Canada, Panama.

Timeline

Recall initiated
2024-06-27
Posted by FDA
2024-09-11
Status

Source: openFDA Device Recall endpoint. Recall record ID #209527. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.