Product

smiths medical portex, Y' PIECE 15MM, REF 100/276/000

FDA product code
BYDCondenser, Heat And Moisture (Artificial Nose)
Device class
Class 1
Medical specialty
Anesthesiology
Affected lot / code info
UDI/DI 35019315011003, Lot Numbers: 4131841, UDI/DI 35019315022405, Lot Numbers: 4122012, 4133227

Why it was recalled

Uncertainty in the seal integrity of the sterile packaging.

Root cause (FDA determination)

Packaging process control

Action the firm took

Smiths medical issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 07/18/2024 via traceable mailing. The notice explained the issue with the products, potential risk and requested the affected devices be destroyed. Distributors were directed to notify their customers by forwarding the notice. Questions - contact smithsmedical.@sedgwick.com.

Recalling firm

Firm
Smiths Medical ASD Inc.
Address
6000 Nathan Ln N, Minneapolis, Minnesota 55442-1690

Distribution

Distribution pattern
Worldwide distribution.

Timeline

Recall initiated
2024-07-18
Posted by FDA
2024-10-23
Status