#209873

Product

ACUSON Maple 1.0 Diagnostic Ultrasound System, REF: 11711750, with software. ACUSON Maple Select 1.0 Diagnostic Ultrasound System, REF: 11711779, with software

FDA product code
IYNSystem, Imaging, Pulsed Doppler, Ultrasonic
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K232617
Affected lot / code info
11711750, UDI-DI: 04056869995809, Software VA10A 11711779, UDI-DI: 04056869995830, Software VA10A

Why it was recalled

If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management

Root cause (FDA determination)

Software design

Action the firm took

On 8/15/2024, correction notices were distributed to customers who were asked to do the following: To avoid potential misdiagnosis by an under-represented volume flow measurement value, use the factory default display configuration of l/min (liters per minute). Review of previous examination results ONLY if a clinical assessment reviewed a volume flow measurement(s) with the unit of measurement configured to display a value in milliliters. Firm's Customer Service Engineer will contact you to schedule a facility visit to update the system or inform you of a remote update when the software update is available. The software update is currently under development and estimated to be available in the summer of 2024. Ensure that all users of ACUSON Juniper and ACUSON Maple systems within your organization, and others who may need to be informed, receive the safety relevant information provided with this notice and take the actions specified herein. If you have further questions, please contact the firm's Ultrasound Service Customer Care Center at 1-800-888-7436.

Recalling firm

Firm
Siemens Medical Solutions USA, Inc.
Address
22010 Se 51st St, Issaquah, Washington 98029-7298

Distribution

Distribution pattern
US: MS, WA, TX, NJ, GA, PN, NY, AL, MN, NE, IN, CT, CA, IL, TN, WI, ND, NM, AR, NC, PR, RI, WY, MD, FL, HI, AZ, SC, MO, OK, VA, ME, KS, MI, LA, IA, OH, SD, NH, KY, AL, OR, MA, GU. OUS: Afghanistan, Albania, Algeria, Andorra, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia-Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Cameroon, Canada, Chile, China, Colombia, Congo, the Democratic Republic of the, Costa Rica, C¿te dIvoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Eritrea, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guinea, French Polynesia, Georgia, Germany, Greece, Guatemala, Guernsey, Guyana, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kosovo, Kuwait, Latvia, Lebanon, Libya, Liechtenstein LI B, Lithuania, Luxembourg, Macedonia, Malawi, Malaysia, Malta, Mexico, Moldova, Rep. of, Mongolia, Montenegro, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic Korea, Reunion, Romania, Ruanda, Russian Federation, San Marino, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad,Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe

Timeline

Recall initiated
2024-08-15
Posted by FDA
2024-09-24
Status

Source: openFDA Device Recall endpoint. Recall record ID #209873. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.