FDA Device Recalls

Recalls /

#210773

Product

Acrobat-i Vacuum Stabilizer System. Model Number C-OM-10000Z

FDA product code
MWSStabilizer, Heart
Device class
Class 1
Medical specialty
Cardiovascular
Affected lot / code info
Model Number C-OM-10000Z. UDI-DI Number: 00607567100008. Batch Numbers: 3000277221, 3000286549, 3000315868, 3000317517, 3000322538, 3000339447, 3000363028, 3000270713, 3000275435, 3000275771, 3000280868, 3000284390, 3000303149, 3000321642, 3000342528, 3000336070, 3000322179, 3000337396, 3000273182, 3000282181, 3000337880, 3000353535, 3000364172, 3000357109, 3000276109, 3000338107, 3000354445, 3000355050, 3000352899, 3000324785, 3000360894, 3000370720, 3000323627, 3000346330, 3000356740, 3000278665, 3000344012, 3000366729, 3000361801, 3000400928, 3000363775, 3000311956, 3000343553, 3000367559, 3000377818, 3000341430, 3000367345, 3000336377, 3000316208, 3000381449, 3000380956, 3000403756, 3000321414, 3000361499, 3000407809, 3000398030, 3000387678, 3000320077, 3000277072, 3000283754, 3000296552, 3000320660, 3000339096, 3000347694, 3000348390, 3000352250, 3000279303, 3000277309, 3000282794, 3000283482, 3000350599, 3000351774, 3000351996, 3000365741, 3000369223, 3000369738, 3000393258, 3000323129, 3000273547, 3000366133, 3000391146, 3000273843, 3000280827, 3000295912, 3000346410, 3000401855, 3000384033, 3000382967, 3000378703, 3000363755, 3000316146, 3000295949, 3000398861, 3000394893, 3000380542, 3000361797, 3000350598, 3000320073, 3000299042, 3000295186, 3000288696, 3000284723, 3000274876, 3000363227, 3000285664.

Why it was recalled

Deviations in the manufacturing process created a compromise in the sterile barrier, in which sterility cannot be assured during the duration of the product's 2-year shelf life.

Root cause (FDA determination)

Process change control

Action the firm took

MAQUET Cardiovascular LLC / Getinge notified consignees via letter FedEx 2-Day Delivery on 10/16/2024. Customers were instructed to forward the notification to all users and notify customers if further distributed, examine inventory and return any affected inventory, and to complete and return the MEDICAL DEVICE REMOVAL - RESPONSE FORM.

Recalling firm

Firm
Maquet Cardiovascular, LLC
Address
45 Barbour Pond Dr, Wayne, New Jersey 07470-2094

Distribution

Distribution pattern
Nationwide distribution. International distribution to Albania, Algeria, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brunei Canada, Croatia, Czech Republic, Denmark, Finland, France and Belarus, Germany, Greece, Hong Kong, India, Indonesia, Italy, Japan, Lebanon, Luxembourg, Malaysia, Netherlands, Oman, Pakistan, Poland, Portugal, Slovenia, South Africa, Spain, Switzerland, the Republic of Korea (South Korea), Turkey, United Arab Emirates, United Kingdom.

Timeline

Recall initiated
2024-10-16
Posted by FDA
2024-11-18
Status

Source: openFDA Device Recall endpoint. Recall record ID #210773. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.