Recalls / —
—#210793
Product
MEDLINE convenience kits labeled as: 1) CVL PACK, REF DYNJ00281O; 2) VASCULAR ACCESS PACK-LF, REF DYNJ0101291F; 3) MAJOR VASCULAR PACK-LF, REF DYNJ0488546K; 4) PEDS CENTRAL LINE PACK, REF DYNJ84762; 5) PACK,CENTRAL LINE, REF DYNJ906907C
- FDA product code
- OFF — Central Venous Catheter Tray
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- Affected lot / code info
- 1) REF DYNJ00281O, UDI/DI 10193489962963 (each) 40193489962964 (case), Lot Numbers: 22KMA417, 22FMD671; 2) REF DYNJ0101291F, UDI/DI 10195327286644 (each) 40195327286645 (case), Lot Numbers: 23FDA396, 23HDA759, 23EDB388, 23LDB909; 3) REF DYNJ0488546K, UDI/DI 10195327505097 (each) 40195327505098 (case), Lot Numbers: 23IBP983; 4) REF DYNJ84762, UDI/DI 10195327413088 (each) 40195327413089 (case), Lot Numbers: 23IMA327, 23IMA727; 5) REF DYNJ906907C, UDI/DI 10195327376819 (each) 40195327376810 (case), Lot Numbers: 23DBM940, 23FBG165, 23IBQ680
Why it was recalled
Medline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.
Root cause (FDA determination)
Under Investigation by firm
Action the firm took
Medline issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 10/11/2024 via mail and email. The notice explained the problem with the Stryker component, potential risk, requested the kits be quarantined pending additional labeling and o respond through their portal: https://recalls.medline.com. Once the response is received, Medline will provide additional labeling to be applied to the affected kits which states the following: "Attention: R-24-219, This kit contains a recalled item, STRYKER SafeAir Smoke Evacuation Pencil, VPN: 0703-047-000, 0703-047-001, 0703-047-002, 0703-046-000, 0703-046-001, 0703-046-002. DO NOT USE THIS COMPONENT. The rest of the kit may be used. Please remove this item only and replace with product from supply. *Do not cover or remove this sticker*". Those further distributing the product were directed to notify their customers.
Recalling firm
- Firm
- MEDLINE INDUSTRIES, LP Northfield
- Address
- 3 Lakes Dr, Northfield, Illinois 60093-2753
Distribution
- Distribution pattern
- US Nationwide distribution.
Timeline
- Recall initiated
- 2024-10-09
- Posted by FDA
- 2024-12-05
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #210793. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.