Recalls / —
—#210870
Product
Atellica CI Analyzer. Catalog Numbers: 10947347.
- FDA product code
- JJE — Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Device class
- Class 1
- Medical specialty
- Clinical Chemistry
- 510(k) numbers
- K222116
- Affected lot / code info
- UDI-DI Number: 630414229560; Catalog Number: 10947347. Software version 1.28.51
Why it was recalled
Potential that the IMT Diluent volume remaining (% remaining) does not decrease as expected on the Atellica CI Analyzer, potentially leading to the IMT Diluent being empty while still displaying that volume is remaining. In this case Sodium (Na), Potassium (K) and Chloride (Cl) test results may be falsely elevated. Quality Control materials demonstrate the same behavior.
Root cause (FDA determination)
Software design
Recalling firm
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Address
- 511 Benedict Ave, Tarrytown, New York 10591-5005
Distribution
- Distribution pattern
- Worldwide distribution - US Nationwide and the countries of Austria, Bangladesh, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Iraq, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Malaysia, Netherlands, Oman, Pakistan, Paraguay, Peru, Philippines, Portugal, Qatar, Republic Korea, Romania, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom Great Britain, Vietnam.
Timeline
- Recall initiated
- 2024-11-04
- Posted by FDA
- 2024-12-04
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #210870. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.