Recalls / —
—#211371
Product
Medline custom medical procedure kits labeled as: 1) CATH LAB PACK-LF, REF DYNJ0382559M; 2) CRANI, REF DYNJ56001B; 3) GENDER AFFIRMING SURGERY, REF DYNJ66252; 4) GENDER AFFIRMING SURGERY, REF DYNJ66252A; 5) MAJOR NEURO PACK-LF, REF DYNJ0578916X.
- FDA product code
- OES — Cardiac Catheterization Kit
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- Affected lot / code info
- 1) REF DYNJ0382559M, UDI/DI 10889942063994 (EA) 10889942063995 (CS), Lot Numbers: 19TBC792; 2) REF DYNJ56001B, UDI/DI 10889942798278 (EA) 40889942798279 (CS), Lot Numbers: 19BDC737, 19FDA310, 19FDB230, 19JDA441; 3) REF DYNJ66252, UDI/DI 10193489324785 (EA) 40193489324786 (CS), Lot numbers: 20DBK065, 20FBN251, 20HBL176, 20HBM692, 21CBP776; 4) REF DYNJ66252A, UDI/DI 10193489887853 (EA) 40193489887854 (CS), Lot numbers: 21IBN151, 21KBO915, 21LBJ105, 21VBB714; 5) REF DYNJ0578916X, UDI/DI 10193489288384 (EA) 40193489288385 (CS), Lot numbers: 22OBL262,20CKA219, 20CKA875, 20EKB076, 20GKA543, 20GKB272, 20IKB497, 20JKB716, 21AKA500, 21BKA195, 21FMD808, 21HMD657, 21IMB172, 21JMH627, 21KMH877, 21LMC161, 22AMA533, 22DMC612, 22DMI233, 22GMB685, 22IMA041, 23AMD730, 23DMA094, 23FMF542, 23IME618, 24AMA230.
Why it was recalled
Custom convenience kits are being recall because the contain non-sterile Codman Surgical Patties & Surgical Strips, which have the potential for endotoxin contamination.
Root cause (FDA determination)
Nonconforming Material/Component
Action the firm took
Medline issued a MEDICAL DEVICE VOLUNTARY RECALL notice on 11/21/2024 to its consignees via US Mail and email. The notice explained the issue, potential risk to health, and requested that the affected product be quarantined. Once the response is received, Medline will provide labels to place on affected inventory, with instructions for the consignee to remove the affected component prior to using the kit.
Recalling firm
- Firm
- MEDLINE INDUSTRIES, LP Northfield
- Address
- 3 Lakes Dr, Northfield, Illinois 60093-2753
Distribution
- Distribution pattern
- Worldwide distribution - US Nationwide and the countries of UAE, and Canada.
Timeline
- Recall initiated
- 2024-11-21
- Posted by FDA
- 2024-12-30
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #211371. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.