FDA Device Recalls

Recalls /

#211931

Product

8MM,MARYLAND BIPOLAR FORCEPS,IS4000 REF 470172

FDA product code
NAYSystem, Surgical, Computer Controlled Instrument
Device class
Class 2
Medical specialty
Gastroenterology, Urology
510(k) numbers
K131861
Affected lot / code info
Model Number: 470172 Part Number: 470172-17 UDI-DI code: 00886874112281 Batch Numbers: K10221204 K10230219 K10230226 K10230302 K10230310 K10230427 K10230518 K10230525 K10230601 K10230608 K10230615 K10230622 K10230629 K10230913 K10231019 K10231123 K10240222 K10240327 K10240404 K10240418 K10240425 K10240502 K10240711 K11231019 K11231116 K11240502 K11240711

Why it was recalled

Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.

Root cause (FDA determination)

Process change control

Action the firm took

On 12/19/2024, the firm send via email an "Urgent: Medical Device Correction" Letter informing customer that Intuitive Surgical has become aware of an increase in complaints regarding frayed or broken cables on some da Vinci X and Xi reusable instruments. Customer are instructed to: 1. Customers can continue using the products in accordance with the user manual. 2. As a reminder, when using da Vinci X and Xi reusable instruments, follow the Inspection Before Use and Warnings listed in the manual provided with your system to inspect for any broken cables. Refer to Figures C and D for examples. In addition, please refer to Appendix B for additional images for detection of grip cable failures. 3. If you observe any failed (frayed or broken) grip cables prior to use, during procedure, or during reprocessing, please stop use of instrument, remove from use and inform Intuitive via the standard complaint process. 4. Inform Intuitive of any Adverse Events*/Serious Incidents** or quality problems concerning the use of the subject device via the standard complaint process. For Question and assistance - " North and South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: customerservice@intusurg.com. " Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) or eucs@intusurg.com " South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) or support.korea@intusurg.com " Japan: 0120-56-5635 or 03-5575-1362 (9 am to 6 pm JST) or csjapan@intusurg.com " India: +1-800-103-6952 (9 am to 6 pm IT) " Taiwan: +0800-86-8181 (9 am to 6 pm CT)

Recalling firm

Firm
Intuitive Surgical, Inc.
Address
1266 Kifer Rd, Sunnyvale, California 94086-5304

Distribution

Distribution pattern
U.S. - Nation-wide. Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming and Puerto Rico. O.U.S: Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Columbia, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, French Guiana, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Luxembourg, Malaysia, Malta, Martinique, Mexico, Monaco, Netherlands, New Zealand, Norway, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, and Vietnam

Timeline

Recall initiated
2024-12-19
Posted by FDA
2025-01-10
Status

Source: openFDA Device Recall endpoint. Recall record ID #211931. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.