Recalls / —
—#212689
Product
Medline procedure kits labeled as: 1) CSMC/ABDOMINAL HYST PACK-LF, Pack Number DYNJ0665086F; 2) MAJOR VAGINAL HARPER PACK-LF, Pack Number DYNJ0161768D; 3) ROBOT LAP TOTAL HYST, Pack Number DYNJ907811A; 4) VAG HYSTE PACK-LF, Pack Number DYNJ0773613J; 5) VAG HYSTERECTOMY PACK-LF, Pack Number DYNJ0549338Q
- FDA product code
- OJF — Hysterectomy Kit
- Device class
- Class 2
- Medical specialty
- Obstetrics/Gynecology
- Affected lot / code info
- 1) Pack Number DYNJ0665086F: UDI/DI 10193489395204 (each) 40193489395205 (case), Lot Number 23JBW085A; 2) Pack Number DYNJ0161768D: UDI/DI 10195327107499 (each) 40195327107490 (case), Lot Number 22CDB491A; 3) Pack Number DYNJ907811A: UDI/DI 10193489984170 (each) 40193489984171 (case), Lot Number 21IBP928Z; 4) Pack Number DYNJ0773613J: UDI/DI 10193489863918 (each) 40193489863919 (case), Lot Number 21GMA889A; 5) Pack Number DYNJ0549338Q: UDI/DI 10195327092139 (each) 40195327092130 (case), Lot Number 23EMD924A
Why it was recalled
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health.
Root cause (FDA determination)
Under Investigation by firm
Action the firm took
Medline Industries LP issued a recall notice to its consignees on 02/11/2025 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. This is an expansion of RES 94583
Recalling firm
- Firm
- MEDLINE INDUSTRIES, LP Northfield
- Address
- 3 Lakes Dr, Northfield, Illinois 60093-2753
Distribution
- Distribution pattern
- Worldwide distribution - US Nationwide and the country of Canada.
Timeline
- Recall initiated
- 2025-02-11
- Posted by FDA
- 2025-03-25
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #212689. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.