Recalls / —
—#212717
Product
GENERAL ANGIO PACK , Model Nos LLAP75G LLAP75G-01 LLAP75G-02 LLAP75G-03
- FDA product code
- OEQ — Angiography/Angioplasty Kit
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- Affected lot / code info
- UDI-DI 00191072219510 Lots 70-050850 70-051285 70-051779 70-051779A 70-052186 70-052669 70-053022
Why it was recalled
ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.
Root cause (FDA determination)
Process control
Action the firm took
The affected distributor was notified via email on 2/16/25, and an updated letter was provided on 2/18/25. The distributor was instructed to notify their end users. Customers were provided with a list of impacted components which are included in the affected trays. Customers were provided with a labeling template and instructed to add warning labels to affected procedure trays to ensure the affected components are discarded.
Recalling firm
- Firm
- American Contract Systems Inc
- Address
- 85 Shaffer Park Dr, Tiffin, Ohio 44883-9290
Distribution
- Distribution pattern
- US Nationwide distribution in the state of OH.
Timeline
- Recall initiated
- 2025-02-06
- Posted by FDA
- 2025-03-14
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #212717. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.