FDA Device Recalls

Recalls /

#212722

Product

URO ROBOTIC PACK , Model No LVUR44C LVUR44C-03

FDA product code
LROGeneral Surgery Tray
Device class
Class 2
Medical specialty
General, Plastic Surgery
Affected lot / code info
UDI-DI 00191072136688 Lots 905241 70-050739 70-051020 70-051480 70-051606 70-051888

Why it was recalled

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Root cause (FDA determination)

Process control

Action the firm took

The affected distributor was notified via email on 2/16/25, and an updated letter was provided on 2/18/25. The distributor was instructed to notify their end users. Customers were provided with a list of impacted components which are included in the affected trays. Customers were provided with a labeling template and instructed to add warning labels to affected procedure trays to ensure the affected components are discarded.

Recalling firm

Firm
American Contract Systems Inc
Address
85 Shaffer Park Dr, Tiffin, Ohio 44883-9290

Distribution

Distribution pattern
US Nationwide distribution in the state of OH.

Timeline

Recall initiated
2025-02-06
Posted by FDA
2025-03-14
Status

Source: openFDA Device Recall endpoint. Recall record ID #212722. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.

— recall: URO ROBOTIC PACK , Model No LVUR44C LVUR44C-03 · FDA Device Recalls