Recalls / —
—#212735
Product
MAJOR VASCULAR PACK , Model No UTMV78X-01 UTMV78Y UTMV78Y-01 UTMV78X-01 UTMV78Y-02
- FDA product code
- OEZ — Cardiovascular Procedure Kit
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- Affected lot / code info
- UDI-DI 00191072205100 00191072217370 Lots 974241 884241 70-051159 70-052011 70-052602 70-053158
Why it was recalled
ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.
Root cause (FDA determination)
Process control
Action the firm took
The affected distributor was notified via email on 2/16/25, and an updated letter was provided on 2/18/25. The distributor was instructed to notify their end users. Customers were provided with a list of impacted components which are included in the affected trays. Customers were provided with a labeling template and instructed to add warning labels to affected procedure trays to ensure the affected components are discarded.
Recalling firm
- Firm
- American Contract Systems Inc
- Address
- 85 Shaffer Park Dr, Tiffin, Ohio 44883-9290
Distribution
- Distribution pattern
- US Nationwide distribution in the state of OH.
Timeline
- Recall initiated
- 2025-02-06
- Posted by FDA
- 2025-03-14
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #212735. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.