FDA Device Recalls

Recalls /

#212778

Product

NexGen LPS Flex, Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis, Model Number 00-5962-042-10

FDA product code
JWHProsthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K042271
Affected lot / code info
UDI-DI: 00889024666214; Lot Numbers: 66789408, 66792792

Why it was recalled

The "Use with plate 7, 8, 9, 10" compatibility statement on top of the label does not match with the compatibility statements "USE WITH PLATE 5, 6" in the translations section of the same label.

Root cause (FDA determination)

Labeling Change Control

Action the firm took

An URGENT MEDICAL DEVICE RECALL notification letter dated 2/27/25 was sent to customers. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with this notice and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Distributor Responsibilities: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form for each return and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form must be returned even if you do not have affected products available to return in your territory. b. For International Returns, request an IRA by emailing zimmerbiometintlirarequests@zimmerbiomet.com c. Include

Recalling firm

Firm
Zimmer, Inc.
Address
1800 W Center St, Warsaw, Indiana 46580-2304

Distribution

Distribution pattern
Worldwide - US Nationwide distribution in the states pf CA, KS, MA, MD, ME, NE, NJ, OR, SD, and TX. The countries of Canada, China, India, Netherlands, Singapore.

Timeline

Recall initiated
2025-02-27
Posted by FDA
2025-03-24
Status

Source: openFDA Device Recall endpoint. Recall record ID #212778. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.