FDA Device Recalls

Recalls /

#213181

Product

Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-10 Software Version: All software versions Product Description: CELL-DYN Ruby Component: No

FDA product code
GKZCounter, Differential Cell
Device class
Class 2
Medical specialty
Hematology
510(k) numbers
K061667
Affected lot / code info
UDI-DI 00380740017170 SNs: 35405BG 36034BG 36259BG 36326BG 36327BG 36328BG 36372BG 36545BG 36556BG 36583BG 36588BG 36589BG 54539BG 55361BG 55454BG 55520BG 55814BG 55828BG 72229BG

Why it was recalled

When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.

Root cause (FDA determination)

Software change control

Action the firm took

On March 18, 2025, the firm notified customers through "Urgent Medical Device Correction" letters delivered via FedEx. The letters described the product, problem and actions to be taken. The Customers were instructed to not use expired reagents as this may compromise test accuracy. Always use reagents within their valid expiration dates for all testing procedures. The operator must confirm the container labeling matches the data on the reagent log entry screen dialog box. Quality Controls should be run after any reagent lot number change, maintenance, component replacement, field service action, software change, or calibration. This should be done in accordance with your laboratory s quality control program and regulatory requirements. Complete and return the Customer Reply Form via email: PMS@abbott.com or fax: 1-800-777-0051. If you have forwarded the product listed above to other laboratories, please inform them of this Product Correction, provide to them a copy of this letter and retain this letter for your laboratory records. If you or any of the health care providers you serve have questions regarding this information, please contact Customer Service at 1-877-4ABBOTT.

Recalling firm

Firm
Abbott Laboratories
Address
100 Abbott Park Rd, Bldg Ap8b, Abbott Park, Illinois 60064-3502

Distribution

Distribution pattern
Worldwide distribution: US (nationwide) to the following states: AR, AZ, CA, CO, FL, GA, IA, IL, KS, LA, MA, MD, MI, MN, MO, MS, ND, NE, NJ, NV, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV, WY and OUS (foreign) to countries of: Afghanistan, Albania, Algeria, Angola, Anguilla, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Croatia, Czech Rebublic, Ecuador, Egypt, El Salvador, Ethiopia, France, Gambia, Gaza & Jericho, Germany, Greece, Guatemala, Guana, Haiti, Honduras, Hong King, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Lebanon, Lithuania, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, New Caledonia, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, SRI Lanka, St. Lucia, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turk & Caicos, Turkey, UAE, Uganda, UK, Ukraine, Uruguay, Uzbekistan, Vietnam, and Yemen.

Timeline

Recall initiated
2025-03-18
Posted by FDA
2025-04-22
Status

Source: openFDA Device Recall endpoint. Recall record ID #213181. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.