Stratus CS Acute Care cTNI TestPak.IVD test for cardiac troponin I in heparinized plasma.

FDA product code
JHXFluorometric Method, Cpk Or Isoenzymes
Device class
Class 2
Medical specialty
Clinical Chemistry
510(k) numbers
K051650
Affected lot / code info
Siemens Materials No. 10445071; UDI-DI: 0405686902174VK; Lot No. 234337002 and forward.

Why it was recalled

Siemens Healthcare Diagnostics Inc., is recalling their Stratus CS Acute Care cTNI TestPak, an IVD test for cardiac troponin I in heparinized plasma by removal. The reason for the recall is an increased occurrence of random non-repeatable false positive cardiac Troponin l(cTnI) results at any point during the TestPaks shelf life when using the Stratus CS cTnI Acute Care Testpak.

Root cause (FDA determination)

Under Investigation by firm

Action the firm took

An Urgent Medical Device Correction notice dated March 2025 was mailed to consignees. The notice instructed consignees to review the provided notification with their medical director to determine the appropriate course of action. Siemens will reimburse users for repeat tests and discarded units associated with this recall; consignees should contact the Customer Care Center for reimbursement. If product was further distributed, Siemens asks that the recall notification be forwarded. The recall notification is to be retained with laboratory records and forwarded to those who use the recalled devices. Consignees with any questions can contact Siemens Healthineers Technical Solutions Center at 977-229-3711.

Recalling firm

Firm
Siemens Healthcare Diagnostics Inc
Address
2 Edgewater Dr, Norwood, Massachusetts 02062-4637

Distribution

Distribution pattern
Worldwide - US Nationwide distribution in the states of CA, IL, KS, MI, MO, NJ, NY, OH, OK, PA, SC & WI. The countries of Austria, Canada, Chile, France, Germany, Iraq, Italy, Malaysia, North Macedonia, Saudi Arabia, Serbia, Slovenia, Spain, Sweden, Switzerland & Vietnam.

Timeline

Recall initiated
2025-03-31
Posted by FDA
2025-04-25