Recalls / —
—#213483
Product
Halyard SUTURE PACK CABG. Cardiovascular Procedure Convenience Kit.
- FDA product code
- OEZ — Cardiovascular Procedure Kit
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- Affected lot / code info
- Model No. SILV800-01; UDI: 10809160450680; Lot Code: Lot: 1642454; Exp: 2/4/2028.
Why it was recalled
An issue with a raw material used to coat some surgical needles that may result in diminished performance and/or removal of some silicone needle coating during use.
Root cause (FDA determination)
Nonconforming Material/Component
Action the firm took
An URGENT: MEDICAL DEVICE RECALL NOTIFICATION dated 3/31/25 was emailed to consignees. The notification indicates that this recall is due to Ethicon having recalled PDS II, PDS Plus, PROLENE, and ATHIBOND EXCEL sutures, which are components in the convenience kits being recalled by the recalling firm. Consignees are being instructed to discontinue use of the Johnson & Johnson Ethicon Sutures contained inside the recalled AVID kits identified in the recall notification and to add the provided labels to affected kits to ensure recalled sutures are discarded. Distributors are asked to forward the provided recall notification to their customers to which product was further distributed. Consignees are asked to return the provided response form via email. Questions regarding this recall event are to be directed to your distributor or sales representative.
Recalling firm
- Firm
- AVID Medical, Inc.
- Address
- 9000 Westmont Dr, Toano, Virginia 23168-9351
Distribution
- Distribution pattern
- US Nationwide distribution in the states of IL & LA.
Timeline
- Recall initiated
- 2025-03-31
- Posted by FDA
- 2025-05-09
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #213483. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.